Partial-volume effect correction in positron emission tomography brain scan image using super-resolution image reconstruction
Authors:
- S TepmongkolSTS Tepmongkol
- C PluempitiwiriyawejCPC Pluempitiwiriyawej
- 1 February 2015
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 88 (1046) , 20140119
Abstract
The partial-volume effect (PVE) is a consequence of limited (i.e. finite) spatial resolution. PVE can lead to quantitative underestimation of activity concentrations in reconstructed images, which may result in misinterpretation of positron emission tomography (PET) scan images, especially in the brain. The PVE becomes significant when the dimensions of a source region are less than two to three times the full width at half maximum spatial resolution of the imaging system. In the present study, the ability of super-resolution (SR) image reconstruction to compensate for PVE in PET was characterized. The ability of SR image reconstruction technique to recover activity concentrations in small structures was evaluated by comparing images before and after image reconstruction in the NEMA/IEC phantom (Washington, DC), in the Hoffman brain phantom and in four human brain subjects (three normal subjects and one atrophic brain subject) in terms of apparent recovery coefficient (ARC) and percentage yield. Both the ARC and percentage yield are improved after SR implementation in NEMA/IEC phantom and Hoffman brain phantom. When tested in normal subjects, SR implementation can improve the intensity and justify SR efficiency to correct PVE. SR algorithm can be used to effectively correct PVE in PET images. The current research focused on brain PET scanning exclusively; future work will extend to whole-body imaging.Keywords
This publication has 18 references indexed in Scilit:
- Comparison between two super‐resolution implementations in PET imagingMedical Physics, 2009
- Partial-Volume Effect in PET Tumor ImagingJournal of Nuclear Medicine, 2007
- Comparative assessment of statistical brain MR image segmentation algorithms and their impact on partial volume correction in PETNeuroImage, 2006
- A multiresolution image based approach for correction of partial volume effects in emission tomographyPhysics in Medicine & Biology, 2006
- Evaluation of anatomy based reconstruction for partial volume correction in brain FDG-PETNeuroImage, 2004
- Robustness of Anatomically Guided Pixel-by-Pixel Algorithms for Partial Volume Effect Correction in Positron Emission TomographyJournal of Cerebral Blood Flow & Metabolism, 1999
- MR-Based Correction of Brain PET Measurements for Heterogeneous Gray Matter Radioactivity DistributionJournal of Cerebral Blood Flow & Metabolism, 1996
- Measurement of Radiotracer Concentration in Brain Gray Matter Using Positron Emission Tomography: MRI-Based Correction for Partial Volume EffectsJournal of Cerebral Blood Flow & Metabolism, 1992
- Improving resolution by image registrationCVGIP: Graphical Models and Image Processing, 1991
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1979