Measurement of the difference in intraocular pressure between the sitting and lying body positions in healthy subjects: direct comparison of the Icare Pro with the Goldmann applanation tonometer, Pneumatonometer and Tonopen XL
Authors:
- 13 January 2014
- journal article
- research article
- Published by Wiley in Clinical & Experimental Ophthalmology
- Vol. 42 (7) , 608-614
Abstract
Change in intraocular pressure is known to occur with body posture. Previously, markedly different estimations of this change were reported. The Icare Pro (Icare; Tiolat Oy, Helsinki, Finland) is designed to measure intraocular pressure with the subject supine. We compared the measurement of postural change in intraocular pressure obtained with this and 3 other tonometers: Goldmann applanation tonometer (CSO, Firenze, Italy), Pneumatonometer (Model 30, Reichert, Depew, NY, USA) and the Tonopen XL (Reichert). Prospective comparative study. Twenty-one healthy subjects. Intraocular pressure was measured in random order with four tonometers, first sitting then after 10 min of lying. Average postural intraocular pressure change (mmHg) and the 95% limits of agreement for measuring postural intraocular pressure change between tonometers. Average postural intraocular pressure change (lying minus sitting, mmHg) was measured highest with the Goldmann applanation tonometer (4.1 ± 2.6) and Pneumatonometer (3.9 ± 2.1), significantly lower with the Tonopen (0.9 ± 1.7) and was negative with the Icare (-0.9 ± 3.3). The 95% limits of agreement for measuring postural intraocular pressure change between Goldmann applanation tonometer and Icare, Pneumatonometer and Tonopen were -3.2-13.3, -6.4-6.9 and -2.8-9.4 mmHg, respectively. Postural change in intraocular pressure was measured similarly, on average, with Goldmann applanation tonometer and Pneumatonometer, and much lower with Tonopen and Icare Pro. There was poor interdevice agreement in measuring this parameter. We observed high intersubject variability, suggesting the clinical importance of this measurement in the diagnosis and management of glaucoma.Keywords
This publication has 32 references indexed in Scilit:
- Effect of Head and Body Position on Intraocular PressureOphthalmology, 2012
- Comparison of iCare tonometer and Goldmann applanation tonometry in normal corneas and in eyes with automated lamellar and penetrating keratoplastyEye, 2011
- The Circadian Curve of Intraocular Pressure: Can We Estimate Its Characteristics during Office Hours?Investigative Ophthalmology & Visual Science, 2009
- Circadian variations in central corneal thickness and intraocular pressure in patients with glaucomaBritish Journal of Ophthalmology, 2006
- Correlation between office and peak nocturnal intraocular pressures in healthy subjects and glaucoma patientsAmerican Journal of Ophthalmology, 2005
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Postural change of IOP in normal persons and in patients with primary wide open-angle glaucoma and low-tension glaucoma.British Journal of Ophthalmology, 1984
- Ocular hypertension and posture.British Journal of Ophthalmology, 1983
- Rapid pneumatic and Mackey-Marg applanation tonometry to evaluate the postural effect on intraocular pressure.British Journal of Ophthalmology, 1976
- POSTURAL VARIATIONS OF THE INTRAOCULAR PRESSURE AS MEASURED WITH THE MACKAY‐MARG TONOMETERActa Ophthalmologica, 1967