Performance of the cobas® influenza A/B assay for rapid PCR-based detection of influenza compared to Prodesse ProFlu+ and viral culture
Authors:
Open Access
- 1 December 2015
- journal article
- Published by Akademiai Kiado Zrt. in European Journal of Microbiology and Immunology
- Vol. 5 (4) , 236-245
Abstract
Rapid and accurate diagnosis of influenza is important for patient management and infection control. We determined the performance of the cobas® Influenza A/B assay, a rapid automated nucleic acid assay performed on the cobas® Liat System for qualitative detection of influenza A and influenza B from nasopharyngeal (NP) swab specimens. Retrospective frozen and prospectively collected NP swabs from patients with signs and symptoms of influenza collected in universal transport medium (UTM) were tested at multiple sites including CLIA-waived sites using the cobas® Influenza A/B assay. Results were compared to the Prodesse ProFlu+ assay and to viral culture. Compared to the Prodesse ProFlu+ Assay, sensitivities of the cobas® Influenza A/B assay for influenza A and B were 97.7 and 98.6%, respectively; specificity was 99.2 and 99.4%. Compared to viral culture, the cobas® Influenza A/B assay showed sensitivities of 97.5 and 96.9% for influenza virus A and B, respectively; specificities were 97.9% for both viruses. Polymerase chain reaction (PCR)/sequencing showed that the majority of viral culture negative but cobas® Influenza A/B positive results were true positive results, indicating that the cobas® Influenza A/B assay has higher sensitivity compared to viral culture. In conclusion, the excellent accuracy, rapid time to result, and remarkable ease of use make the cobas® Influenza A/B nucleic acid assay for use on the cobas® Liat System a highly suitable point-of-care solution for the management of patients with suspected influenza A and B infection.Keywords
This publication has 38 references indexed in Scilit:
- Performance of the Molecular Alere i Influenza A&B Test Compared to That of the Xpert Flu A/B AssayJournal of Clinical Microbiology, 2015
- Comparison of the Biofire FilmArray RP, Genmark eSensor RVP, Luminex xTAG RVPv1, and Luminex xTAG RVP Fast Multiplex Assays for Detection of Respiratory VirusesJournal of Clinical Microbiology, 2013
- Clinical Accuracy of a PLEX-ID Flu Device for Simultaneous Detection and Identification of Influenza Viruses A and BJournal of Clinical Microbiology, 2013
- How accurate are rapid influenza diagnostic tests?Expert Review of Anti-infective Therapy, 2012
- Clinical Validation of a Point-of-Care Multiplexed In Vitro Immunoassay Using Monoclonal Antibodies (the MSD Influenza Test) in Four Hospitals in VietnamJournal of Clinical Microbiology, 2012
- Comparison of the Idaho Technology FilmArray System to Real-Time PCR for Detection of Respiratory Pathogens in ChildrenJournal of Clinical Microbiology, 2012
- VereFlu™: an integrated multiplex RT-PCR and microarray assay for rapid detection and identification of human influenza A and B viruses using lab-on-chip technologyArchives of Virology, 2011
- Rapid Diagnostic Testing for Seasonal Influenza: An Evidence-Based Review and Comparison with Unaided Clinical DiagnosisThe Journal of Emergency Medicine, 2010
- Performance of six influenza rapid tests in detecting human influenza in clinical specimensJournal of Clinical Virology, 2007
- Large-scale sequencing of human influenza reveals the dynamic nature of viral genome evolutionNature, 2005