A microfabricated amorphous silicon nitride membrane-based nanocalorimeter is proposed to be suitable for an x-ray transparent sample platform with low power heating and built-in temperature sensing. In this work, thermal characterization in both air and vacuum are analyzed experimentally and via simulation. Infrared microscopy and thermoreflectance microscopy are used for thermal imaging of the sample area in air. While a reasonably large isothermal area is found on the sample area, the temperature homogeneity of the entire sample area is low, limiting use of the device as a heater stage in air or other gases. A simulation model that includes conduction, as well as radiation and convection heat loss, is presented with radiation and convection parameters determined experimentally. Simulated temperature distributions show that the homogeneity can be improved by using a thicker thermal conduction layer or reducing the pressure of the gas in the environment but neither are good solutions for the proposed use. A new simple design that has improved temperature homogeneity and a larger isothermal area while maintaining a thin thermal conduction layer is proposed and fabricated. This new design enables applications in transmission x-ray microscopes and spectroscopy setups at atmospheric pressure.Keywords
Funding Information
- U.S. Department of Energy (DE-AC02-05CH11231)
This publication has 13 references indexed in Scilit:
- CCD-based thermoreflectance microscopy: principles and applicationsJournal of Physics D: Applied Physics, 2009
- Thin film nanocalorimeter for heat capacity measurements of 30 nm filmsReview of Scientific Instruments, 2009
- Microscale and Nanoscale Thermal Characterization TechniquesJournal of Electronic Packaging, 2008
- Nanoscale chemical imaging of a working catalyst by scanning transmission X-ray microscopyNature, 2008
- Thermodynamic measurements of submilligram bulk samples using a membrane-based “calorimeter on a chip”Review of Scientific Instruments, 2008
- In situ Transmission Electron Microscopy Studies Enabled by Microelectromechanical System TechnologyJournal of Materials Research, 2005
- Si-N membrane-based microcalorimetry: Heat capacity and thermal conductivity of thin filmsThermochimica Acta, 2005
- Numerical simulation of the heat transfer in amorphous silicon nitride membrane-based microcalorimetersReview of Scientific Instruments, 2003
- True temperature measurements on microscopic semiconductor targets1999
- Thin film microcalorimeter for heat capacity measurements from 1.5 to 800 KReview of Scientific Instruments, 1994