The Dk Project: An Interlaboratory Comparison of DK/L Measurements

Authors:
  • JNJulia Newton-Howes
    Julia Newton-Howes
  • LWLynn Winterton
    Lynn Winterton
  • IFIrving Fatt
    Irving Fatt
  • HHHikaru Hamano
    Hikaru Hamano
  • DHDonna La Hood
    Donna La Hood
  • NBNoel A Brennan
    Noel A Brennan
  • NENathan Efron
    Nathan Efron

Abstract
The oxygen transmissibilities (Dk/L) of a set of 48 contact lenses made from 8 different materials were measured by 4 laboratories. The L/Dk measurements from each laboratory were compared and correlated. Samples which were not masked with a fixed front surface aperture during measurement were corrected for edge effects. This paper shows that provided L/Dk is calculated for each lens using the same technique and Dk is derived using a graphical method of calculation, similar results can be obtained by all laboratories. However, the agreement was less good for materials of Dk greater than 70 x 10(-11) (cm2/s) (ml O2/ml x mm Hg).