A High-Throughput MEMS-Based Differential Scanning Calorimeter for Direct Thermal Characterization of Antibodies
Open Access
- 16 June 2022
- journal article
- research article
- Published by MDPI AG in Biosensors
- Vol. 12 (6) , 422
Abstract
Calorimeters, which can be used for rapid thermal characterization of biomolecules, are getting intense attention in drug development. This paper presents a novel MEMS-based differential scanning calorimeter (DSC) for direct thermal characterization of protein samples. The DSC consisted of a pair of temperature sensors made by vanadium oxide (VOx) film with a temperature coefficient of resistivity of −0.025/K at 300 K, a microfluidic device with high thermal insulation (2.8 K/mW), and a Peltier heater for linear temperature scanning. The DSC exhibited high sensitivity (6.1 µV/µW), low noise (0.4 µW), high scanning rate (45 K/min), and low sample consumption volume (0.63 µL). The MEMS DSC was verified by measuring the temperature-induced denaturation of lysozyme at different pH, and then used to study the thermal stability of a monoclonal antibody (mAb), an antigen-binding fragment (Fab), and a dual variable domain immunoglobulin (DVD-Ig) at pH = 6. The results showed that lysozyme is a stable protein in the pH range of 4.0–8.0. The protein stability study revealed that the transition temperatures of the intact Fab fragment, mAb, and DVD proteins were comparable with conformational stability results obtained using conventional commercial DSC. These studies demonstrated that the MEMS DSC is an effective tool for directly understanding the thermal stability of antibodies in a high-throughput and low-cost manner compared to conventional calorimeters.Keywords
Funding Information
- National Science Foundation (IDBR #1530508, #98467)
- AbbVie Inc. (IDBR #1530508, #98467)
This publication has 33 references indexed in Scilit:
- Design, performance and analysis of thermal lag of the UFS1 twin-calorimeter chip for fast scanning calorimetry using the Mettler-Toledo Flash DSC 1Thermochimica Acta, 2011
- Temperature calibration and electrical characterization of the differential scanning calorimeter chip UFS1 for the Mettler-Toledo Flash DSC 1Thermochimica Acta, 2011
- The Flash DSC 1, a power compensation twin-type, chip-based fast scanning calorimeter (FSC): First findings on polymersThermochimica Acta, 2011
- Miniaturized calorimeter for thermal screening of energetic materialsMicroelectronics Journal, 2010
- Fast scanning power compensated differential scanning nano-calorimeter: 2. Heat capacity analysisThermochimica Acta, 2010
- Therapeutic antibodies: past, present and futureNature Reviews Immunology, 2010
- The Design and Characterization of Oligospecific Antibodies for Simultaneous Targeting of Multiple Disease MediatorsJournal of Molecular Biology, 2009
- Use of Microcalorimetry and Its Correlation with Size Exclusion Chromatography for Rapid Screening of the Physical Stability of Large Pharmaceutical Proteins in SolutionPharmaceutical Development and Technology, 2007
- Domain Interactions in the Fab Fragment: A Comparative Evaluation of the Single-chain Fv and Fab Format Engineered with Variable Domains of Different StabilityJournal of Molecular Biology, 2005
- Influence of experimental variables on curves in differential scanning calorimetry. Part II. Effects on baseline-related characteristicsThermochimica Acta, 1981