A MEMS Convective Accelerometer Equipped with On-Chip Facilities for Sensitivity Electrical Calibration

Authors:
  • ARAhmed Amine Rekik
    Ahmed Amine Rekik
  • FAFlorence Azaïs
    Florence Azaïs
  • NDNorbert Dumas
    Norbert Dumas
  • FMFrederick Mailly
    Frederick Mailly

Abstract
In this paper, we propose an alternate electrical-only strategy to perform both test and calibration of MEMS convective accelerometers' sensitivity. The idea is calibrate device sensitivity through the adjustment of the power dissipated in the heater element. For this, the system is equipped with an on-chip programmable digital pulse modulated generator and an iterative search procedure based on simple impedance measurements is implemented. The method is evaluated through Monte-Carlo simulations considering typical process variations and results are presented to demonstrate the potentialities of the technique.

This publication has 2 references indexed in Scilit: