Protocol optimization for functional cardiac CT imaging using noise emulation in the raw data domain
Authors:
- 16 May 2024
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 51 (7) , 4622-4634
Abstract
Background: Four‐dimensional (4D) wide coverage computed tomography (CT) is an effective imaging modality for measuring the mechanical function of the myocardium. However, repeated CT measurement across a number of heartbeats is still a concern.Purpose: A projection‐domain noise emulation method is presented to generate accurate low‐dose (mA modulated) 4D cardiac CT scans from high‐dose scans, enabling protocol optimization to deliver sufficient image quality for functional cardiac analysis while using a dose level that is as low as reasonably achievable (ALARA).Methods: Given a targeted low‐dose mA modulation curve, the proposed noise emulation method injects both quantum and electronic noise of proper magnitude and correlation to the high‐dose data in projection domain. A spatially varying (i.e., channel‐dependent) detector gain term as well as its calibration method were proposed to further improve the noise emulation accuracy. To determine the ALARA dose threshold, a straightforward projection domain image quality (IQ) metric was proposed that is based on the number of projection rays that do not fall under the non‐linear region of the detector response. Experiments were performed to validate the noise emulation method with both phantom and clinical data in terms of visual similarity, contrast‐to‐noise ratio (CNR), and noise‐power spectrum (NPS).Results: For both phantom and clinical data, the low‐dose emulated images exhibited similar noise magnitude (CNR difference within 2%), artifacts, and texture to that of the real low‐dose images. The proposed channel‐dependent detector gain term resulted in additional increase in emulation accuracy. Using the proposed IQ metric, recommended kVp and mA settings were calculated for low dose 4D Cardiac CT acquisitions for patients of different sizes.Conclusions: A detailed method to estimate system‐dependent parameters for a raw‐data based low dose emulation framework was described. The method produced realistic noise levels, artifacts, and texture with phantom and clinical studies. The proposed low‐dose emulation method can be used to prospectively select patient‐specific minimal‐dose protocols for functional cardiac CT.Keywords
Funding Information
- National Institutes of Health (R01 HL14678)
This publication has 25 references indexed in Scilit:
- Four‐dimensional computed tomography of the left ventricle, Part I: Motion artifact reductionMedical Physics, 2022
- Four‐dimensional computed tomography of the left ventricle, Part II: Estimation of mechanical activation timesMedical Physics, 2022
- Theory, method, and test tools for determination of 3D MTF characteristics in cone‐beam CTMedical Physics, 2021
- Image based simulation of the low dose computed tomography images suggests 13 mAs 120 kV suitability for non-syndromic craniosynostosis diagnosis without iterative reconstruction algorithmsEuropean Journal of Radiology, 2018
- Low‐dose cone‐beam CT via raw counts domain low‐signal correction schemes: Performance assessment and task‐based parameter optimization (Part I: Assessment of spatial resolution and noise performance)Medical Physics, 2018
- Low‐dose CT for the detection and classification of metastatic liver lesions: Results of the 2016 Low Dose CT Grand ChallengeMedical Physics, 2017
- Low-Dose CT With a Residual Encoder-Decoder Convolutional Neural NetworkIEEE Transactions on Medical Imaging, 2017
- Deriving depth‐dependent light escape efficiency and optical Swank factor from measured pulse height spectra of scintillatorsMedical Physics, 2016
- Development and Validation of a Practical Lower-Dose-Simulation Tool for Optimizing Computed Tomography Scan ProtocolsJournal of Computer Assisted Tomography, 2012
- Synthetic CT noise emulation in the raw data domain2010