Feasibility of Crosslinked Acrylic Shape Memory Polymer for a Thrombectomy Device
Open Access
- 25 February 2014
- journal article
- research article
- Published by Hindawi Limited in Smart Materials Research
- Vol. 2014, 1-12
Abstract
Purpose. To evaluate the feasibility of utilizing a system of SMP acrylates for a thrombectomy device by determining an optimal crosslink density that provides both adequate recovery stress for blood clot removal and sufficient strain capacity to enable catheter delivery. Methods. Four thermoset acrylic copolymers containing benzyl methacrylate (BzMA) and bisphenol A ethoxylate diacrylate (Mn∼512, BPA) were designed with differing thermomechanical properties. Finite element analysis (FEA) was performed to ensure that the materials were able to undergo the strains imposed by crimping, and fabricated devices were subjected to force-monitored crimping, constrained recovery, and bench-top thrombectomy. Results. Devices with 25 and 35 mole% BPA exhibited the highest recovery stress and the highest brittle response as they broke upon constrained recovery. On the contrary, the 15 mole% BPA devices endured all testing and their recovery stress (5kPa) enabled successful bench-top thrombectomy in 2/3 times, compared to 0/3 for the devices with the lowest BPA content. Conclusion. While the 15 mole% BPA devices provided the best trade-off between device integrity and performance, other SMP systems that offer recovery stresses above 5kPa without increasing brittleness to the point of causing device failure would be more suitable for this application.Keywords
This publication has 33 references indexed in Scilit:
- The effect of free radical inhibitor on the sensitized radiation crosslinking and thermal processing stabilization of polyurethane shape memory polymersRadiation Physics and Chemistry, 2013
- Porous Shape-Memory PolymersPolymer Reviews, 2013
- Electron Beam Crosslinked Polyurethane Shape Memory Polymers with Tunable Mechanical PropertiesMacromolecular Chemistry and Physics, 2012
- Ultra low density and highly crosslinked biocompatible shape memory polyurethane foamsJournal of Polymer Science Part B: Polymer Physics, 2012
- The Limitations of Thrombectomy with Solitaire™ AB as First-line Treatment in Acute Ischemic Stroke: A Single Center ExperienceJournal of Cerebrovascular and Endovascular Neurosurgery, 2012
- Estimation of aneurysm wall stresses created by treatment with a shape memory polymer foam deviceBiomechanics and Modeling in Mechanobiology, 2011
- Post‐polymerization crosslinked polyurethane shape memory polymersJournal of Applied Polymer Science, 2011
- Structure–property relationships in photopolymerizable polymer networks: Effect of composition on the crosslinked structure and resulting thermomechanical properties of a (meth)acrylate‐based systemJournal of Applied Polymer Science, 2008
- Survey of Emergency Physicians About Recombinant Tissue Plasminogen Activator for Acute Ischemic StrokeAnnals of Emergency Medicine, 2005
- Ultrasonic measurement of the elastic modulus of the common carotid artery. The Atherosclerosis Risk in Communities (ARIC) Study.Stroke, 1992