Optical dialysis adequacy sensor: contribution of chromophores to the ultra violet absorbance in the spent dialysate
- 1 August 2006
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE) in 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
- Vol. 2006 (1557170X) , 807-810
Abstract
Several on-line methods have been developed to standardize the assessment of dialysis adequacy. Earlier studies have demonstrated that on-line monitoring of total ultra violet (UV) absorbance in spent dialysate can be utilized to follow continuously a single hemodialysis session. The aim of this study was to investigate the contribution of different compounds, acting as chromophores, to the UV-absorbance in the spent dialysate in order to explain origin of the cumulative and integrated UV-absorbance measured by the optical dialysis adequacy sensor. Four uremic patients, during 12 hemodialysis treatments, were followed by the optical dialysis adequacy sensor using the wavelength of 280 mn. The dialysate samples were taken and analyzed using reversed phase high performance liquid chromatography (HPLC). The total number of detected peaks from the HPLC gradient separation profiles measured at the wavelength 280 nm for the samples collected 10 mm after the start of hemodialysis (MeanplusmnSD) was 38plusmn6. The relative contribution from the area of 10 main peaks to the total area of all detected peaks in percentage was 91.01plusmn2.52 %. The optical dialysis adequacy sensor provides continuous, on-line hemodialysis measurements and may immediately identify and alert to any deviations in the dialysis. Our study indicates that there exists a number of prevalent compounds that are the main cause of the cumulative and integrated UV- absorbanceKeywords
This publication has 12 references indexed in Scilit:
- Haemodialysis with on-line monitoring equipment: tools or toys?Nephrology Dialysis Transplantation, 2005
- Developments in online monitoring of haemodialysis patients: towards global assessment of dialysis adequacyCurrent Opinion in Nephrology and Hypertension, 2003
- Estimation of delivered dialysis dose by on-line monitoring of the ultraviolet absorbance in the spent dialysateAmerican Journal of Kidney Diseases, 2003
- On-line monitoring of solutes in dialysate using wavelength-dependent absorption of ultraviolet radiationMedical & Biological Engineering & Computing, 2003
- Review on uremic toxins: Classification, concentration, and interindividual variabilityKidney International, 2003
- Assessment of the Length of Each Hemodialysis Session by On-Line Dialysate Urea MonitoringNephron, 2001
- Redesigning the Map of Uremic Toxins2001
- Whither goest Kt/V?Kidney International, 2000
- Mass spectrometry in the search for uremic toxinsMass Spectrometry Reviews, 1997
- Isolation, structure and biological activity of the Trp-containing pentapeptide from uremic fluidBiochemical and Biophysical Research Communications, 1979