Routine Vascular Control Using the Hem-o-Lok Clip in Laparoscopic Nephrectomy: Animal Study and Clinical Application

Authors:

Abstract
Background and Purpose: During laparoscopic nephrectomy (LPN), a stapling device is often used for vascular control, especially of the renal vein. Herein, we report our experience using a polymer clip (Hem-o-lok) for routine control of the vessels during LPN in the animal and clinical setting. Patients and Methods: Fifty ablative and fifteen live-donor nephrectomies were performed in domestic pigs. Hem-o-lok clips (10 mm; Weck Closure System, Research Triangle Park, NC) were routinely used for vascular control. In addition, from January 2001 to July 2002, 46 patients underwent hand-assisted laparoscopic (HAL) (N = 40) or laparoscopic (N = 6) nephrectomy for renal disease or donor nephrectomy. Venous control was achieved solely by the Hem-o-lok clips where at least two clips were applied on the patient side. Arterial control was obtained by the Hem-o-lok clips either alone or in combination with the metal clips. The technical difficulty in obtaining vascular control, transfusion requirement, and clinical outcome were evaluated. Results: In the animal study (total 65 nephrectomies), individual vascular control was obtained by the Hemo-lok clip in all cases except two, where vascular injury during dissection necessitated endoscopic stapling of renal hilum or open conversion. The warm ischemic time for animal donor kidney harvest was uniformly Conclusion: The Hem-o-lok is a reliable and economical device for vascular control in laparoscopic renal surgery.