Advanced experimental application of a digital signal processor in high resolution x-ray spectroscopy
Authors:
- 1 November 1995
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 66 (11) , 5381-5382
Abstract
‘‘Optimum’’ digital filtering of the digitized pulses obtained from a radiation detector had been shown to improve the energy resolution with respect to suboptimum analog processing [Rev. Sci. Instrum. 66, 975 (1995)]. This note shows that the improvement does not vanish in very high resolution systems, as it depends on the kind and relative amount of the noises present and not on their absolute value. Using one of the best available silicon detectors, the electronic noise has again been lowered by 18% from 9.6 to 7.9 electrons rms.Keywords
This publication has 7 references indexed in Scilit:
- A digital system for ‘‘optimum’’ resolution in x-ray spectroscopyReview of Scientific Instruments, 1995
- An optimum digital signal processing for radiation spectroscopyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1994
- A method for the determination of the noise parameters in preamplifying systems for semiconductor radiation detectorsReview of Scientific Instruments, 1993
- Sampling and optimum data processing of detector signalsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1992
- Semiconductor drift detectors for high resolution X-ray spectroscopySensors and Actuators A: Physical, 1992
- Optimum filters for detector charge measurements in presence of noiseNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1990
- Semiconductor drift chambers for position and energy measurementsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1985