Study of ExacTrac X‐ray 6D IGRT setup uncertainty for marker‐based prostate IMRT treatment
Authors:
Open Access
- 10 May 2012
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 13 (3) , 35-42
Abstract
Novalis Tx ExacTrac X-ray system has the 6D adjustment ability for patient setup. Limited studies exist about the setup uncertainty with ExacTrac X-ray system for IMRT prostate treatment with fiducial markers implanted. The purpose of this study is to investigate the marker-based prostate IMRT treatment setup uncertainty using ExacTrac 6D IGRT ability for patient setup. Forty-three patients with prostate cancers and markers implanted have been treated on the Novalis Tx machine. The ExacTrac X-ray system has been used for the patient pretreatment setup and intratreatment verification. In total, the shifts data for 1261 fractions and 3504 correction times (the numbers of X-ray images were taken from tube 1 and tube 2) have been analyzed. The setup uncertainty has been separated into uncertainties in 6D. Marker matching uncertainty was also analyzed. Correction frequency probability density function was plotted, and the radiation dose for imaging was calculated. The minimum, average, and maximum translation shifts were: , , and , respectively, in the lateral direction; , , and , respectively, in the longitudinal direction; and , , and , respectively, in the vertical direction. The minimum, average, and maximum rotation shifts were: , , and , respectively, along lateral direction; , , and , respectively, along longitudinal direction; and , , and , respectively, along vertical direction. On average, each patient had three correction times during one fraction treatment. The radiation dose is about 3 mSv per fraction. With the ExacTrac 6D X-ray system, the prostate IMRT treatment with marker implanted can achieve less than 2 mm setup uncertainty in translations, and less than 0.25° in rotations as overall interfraction mean error. The imaging dose is less than kV (CBCT) for setup verification. PACS number: 87.55.km, 87.55.Qr, 87.56.bd
Keywords
This publication has 12 references indexed in Scilit:
- Intra-fraction setup variability: IR optical localization vs. X-ray imaging in a hypofractionated patient populationRadiation Oncology, 2011
- The geometric accuracy of frameless stereotactic radiosurgery using a 6D robotic couch systemPhysics in Medicine & Biology, 2009
- Quelle radiothérapie guidée par l’image pour quels patients? Expérience concomitante de l’utilisation de trois dispositifs d’imagerie de repositionnement dans le cas du cancer de la prostateCancer/Radiothérapie, 2009
- Setup Variations and Tumor Margins Using the ExacTrac X-Ray 6D in Prostate RadiotherapyIFMBE Proceedings, 2009
- Simulation of intrafraction motion and overall geometric accuracy of a frameless intracranial radiosurgery processJournal of Applied Clinical Medical Physics, 2008
- Use of the BrainLAB ExacTrac X-Ray 6D System in Image-Guided RadiotherapyMedical Dosimetry, 2008
- NOVALIS FRAMELESS IMAGE-GUIDED NONINVASIVE RADIOSURGERYNeurosurgery, 2008
- Accurate patient dosimetry of kilovoltage cone‐beam CT in radiation therapyMedical Physics, 2008
- Prostate Planning Treatment Volume Margin Calculation Based on the ExacTrac X-Ray 6D Image-Guided System: Margins for Various Clinical ImplementationsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Assessment of secondary patient motion induced by automated couch movement during on-line 6 dimensional repositioning in prostate cancer treatmentRadiotherapy and Oncology, 2007