Machine learning-based patient specific prompt-gamma dose monitoring in proton therapy
Open Access
- 14 June 2013
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 58 (13) , 4563-4577
Abstract
Online dose monitoring in proton therapy is currently being investigated with prompt-gamma (PG) devices. PG emission was shown to be correlated with dose deposition. This relationship is mostly unknown under real conditions. We propose a machine learning approach based on simulations to create optimized treatment-specific classifiers that detect discrepancies between planned and delivered dose. Simulations were performed with the Monte-Carlo platform Gate/Geant4 for a spot-scanning proton therapy treatment and a PG camera prototype currently under investigation. The method first builds a learning set of perturbed situations corresponding to a range of patient translation. This set is then used to train a combined classifier using distal falloff and registered correlation measures. Classifier performances were evaluated using receiver operating characteristic curves and maximum associated specificity and sensitivity. A leave-one-out study showed that it is possible to detect discrepancies of 5 mm with specificity and sensitivity of 85% whereas using only distal falloff decreases the sensitivity down to 77% on the same data set. The proposed method could help to evaluate performance and to optimize the design of PG monitoring devices. It is generic: other learning sets of deviations, other measures and other types of classifiers could be studied to potentially reach better performance. At the moment, the main limitation lies in the computation time needed to perform the simulations.Keywords
This publication has 31 references indexed in Scilit:
- Interaction vertex imaging (IVI) for carbon ion therapy monitoring: a feasibility studyPhysics in Medicine & Biology, 2012
- Real-time prompt gamma monitoring in spot-scanning proton therapy using imaging through a knife-edge-shaped slitPhysics in Medicine & Biology, 2011
- A Monte Carlo pencil beam scanning model for proton treatment plan simulation using GATE/GEANT4Physics in Medicine & Biology, 2011
- Extension and validation of an analytical model forin vivoPET verification of proton therapy—a phantom and clinical studyPhysics in Medicine & Biology, 2011
- The reliability of proton-nuclear interaction cross-section data to predict proton-induced PET images in proton therapyPhysics in Medicine & Biology, 2011
- Dynamic Partitioning of GATE Monte-Carlo Simulations on EGEEJournal of Grid Computing, 2010
- A Tracking Compton-Scattering Imaging System for Hadron Therapy MonitoringIEEE Transactions on Nuclear Science, 2010
- Charged particles in radiation oncologyNature Reviews Clinical Oncology, 2009
- Dosimetric Comparison of Three Different Involved Nodal Irradiation Techniques for Stage II Hodgkin's Lymphoma Patients: Conventional Radiotherapy, Intensity-Modulated Radiotherapy, and Three-Dimensional Proton RadiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- On the clinical spatial resolution achievable with protons and heavier charged particle radiotherapy beamsPhysics in Medicine & Biology, 2009