Development of a Protective Device to Prevent Damage to an Endotracheal Tube Caused by the Eruption of Deciduous Teeth
Open Access
Abstract
In patients undergoing endotracheal intubation and mechanical ventilation, damage to the endotracheal tube or inflation line is a potential risk with life-threatening consequences. Although various devices and bite blocks have been developed to secure the endotracheal tube and prevent its dislodgement or damage, they are not suitable for children or newborns. We describe a 1-year-and-2-month-old girl who required long-term endotracheal intubation and mechanical ventilation. The eruption of the lower deciduous incisors caused damage to the inflation line. A mouthguard was not appropriate because her teeth had erupted only slightly, and the retention was insufficient. After collaborating with a dental technician to fabricate a protective device, we were able to provide the optimal treatment while minimizing invasiveness to the patient. It is necessary to collaborate with other professionals as a team to select the optimal medical care tailored to each patient.Keywords
This publication has 9 references indexed in Scilit:
- Clinical application and evaluation of a new type of tracheal catheter fixation beltNursing Open, 2022
- The effect of adhesive tape versus endotracheal tube fastener in critically ill adults: the endotracheal tube securement (ETTS) randomized controlled trialCritical Care, 2019
- Risk factors and outcomes of tracheostomy after prolonged mechanical ventilation in pediatric patients with heart diseaseProgress in Pediatric Cardiology, 2018
- Impact of the introduction of an endotracheal tube attachment device on the incidence and severity of oral pressure injuries in the intensive care unit: a retrospective observational studyBMC Nursing, 2018
- Tracheostomy in Infants and ChildrenRespiratory Care, 2017
- [Anchor Fast endotracheal tube securing device for a pediatric patient during therapeutic hypothermia].2014
- Endotracheal Tube Cuff LeaksAnesthesia & Analgesia, 2013
- Praxis der Fixierung des EndotrachealtubusMedizinische Klinik - Intensivmedizin und Notfallmedizin, 2013
- Custom Built Bite-Block Endotracheal Tube Holder for Prolonged IntubationAnaesthesia and Intensive Care, 2004