Evaluation of EnSeal®, an adaptive bipolar electrosurgical tissue-sealing device
Authors:
- 1 March 2012
- journal article
- Published by Akademiai Kiado Zrt. in Acta Veterinaria Hungarica
- Vol. 60 (1) , 27-40
Abstract
Relatively few, and inconsistent, data are available in the literature about the properties of EnSeal®, an electrosurgical tissue-sealing device. For this reason, we conducted control safety tests on experimental pigs. The mean burst pressure of sealed vessels (2–7 mm in diameter) proved to be 873.89 ± 120.57 mmHg (n = 60). Surface temperature increased to 69.25 ± 0.98 °C in average (n = 22). The mean diameter of the collateral microscopic thermal injury zone was found to be 0.28 ± 0.04 mm, and it did not show significant differences among the groups of tissues studied (n = 183). During our studies, the device worked reliably and met the relevant requirements in all cases. It can be established that EnSeal® enables high-safety clinical interventions at high blood pressure values, in different tissues and even at sites adjacent to heat-sensitive tissues, and thus it paves the way for new operative solutions in both human and veterinary surgery. In our opinion, the discrepancies between data reported in the literature arise from differences in the design of studies and in the designated limit values. To ensure standardisation, we recommend the use of the nitroblue-tetrazolium chloride/lactate dehydrogenase (NBTC/LDH) enzyme histochemical technique for studying thermal injury induced by the different performance levels and application times of devices operating with electromagnetic energy.Keywords
This publication has 11 references indexed in Scilit:
- Electromagnetic Energy Sources in SurgeryVeterinary Surgery, 2010
- Application of Ice Cold Irrigation During Vascular Pedicle Control of Robot-Assisted Radical Prostatectomy: EnSeal Instrument Cooling to Reduce Collateral Thermal Tissue DamageJournal of Endourology, 2010
- Comparative Study of In Vivo Lymphatic Sealing Capability of the Porcine Thoracic Duct Using Laparoscopic Dissection DevicesJournal of Urology, 2009
- Prospective Comparison of Four Laparoscopic Vessel Ligation DevicesJournal of Endourology, 2008
- Comparison of blood vessel sealing among new electrosurgical and ultrasonic devicesSurgical Endoscopy, 2008
- Comparison of four energy-based vascular sealing and cutting instruments: A porcine modelSurgical Endoscopy, 2007
- Comparison of Three Techniques for Ovarian Pedicle Hemostasis During Laparoscopic‐Assisted OvariohysterectomyVeterinary Surgery, 2007
- Histopathological Effects of Bipolar Vessel Sealing Devices on Liver Parenchyma and Comparison with Suture Method: An Experimental StudyEuropean Surgical Research, 2007
- 105: A Preliminary Comparison of Mechanical Compression amongst Three Electrosurgical DevicesJournal of Minimally Invasive Gynecology, 2005
- Mapping the thermal gradient of a new radiofrequency bipolar vessel sealing device, EnSeal, using real-time thermographyThe Journal of the American Association of Gynecologic Laparoscopists, 2004