Pacing Threshold Testing Induced Ventricular Fibrillation Following Acute Rate Control of Atrial Fibrillation
- 23 November 2009
- journal article
- case report
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 20 (12) , 1405-1407
Abstract
Background: A properly placed stimulus on the T‐wave during ventricular repolarization can result in ventricular fibrillation (VF). Initiation of VF with pacing on T‐wave is a rare event with a few reported cases in the literature. We present a unique case of induced VF attributed to a pacing stimulus on T‐wave during ventricular pacing threshold testing of a permanent pacemaker. Case Report: A 64‐year‐old woman with persistent atrial fibrillation (AF) and a permanent pacemaker for tachycardia–bradycardia syndrome presented with symptomatic AF with rapid ventricular response. Acute rate control was achieved with intravenous diltiazem. During ventricular pacing threshold testing, noncapture occurred followed by a pacing spike on T‐wave initiating VF. Cardiopulmonary resuscitation and defibrillation converted the rhythm to rate‐controlled AF. An acute prolongation of the QT was noted and normalized within 12 hours. No antiarrhythmic medications were used. Postevent laboratory values were within normal limits. She was free of ischemia and an echocardiogram revealed normal left ventricular function. She recovered from the event and was discharged with rate‐controlled AF. No further pacing‐induced arrhythmias have occurred during follow‐up pacemaker interrogation and 12‐lead electrocardiograms continued to show normal QT intervals. Conclusion: Pacemaker‐induced VF is an extraordinarily rare complication of cardiac pacing. Alterations in ventricular repolarization with rapid slowing of the heart rate demonstrated by acute prolongation of QT intervals may play a role. This report should alert physicians to the possibility of QT prolongation and an increased risk of ventricular arrhythmias following acute rate control of AF.Keywords
This publication has 16 references indexed in Scilit:
- A Rate-Independent Method of Assessing QT-RR Slope Following Conversion of Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2007
- Pacemaker induced ventricular fibrillation in coronary care unitsBMJ, 2004
- Ventricular Fibrillation and Sudden Death After Radiofrequency Catheter Ablation of the Atrioventricular JunctionPacing and Clinical Electrophysiology, 1997
- Induction of Ventricular Tachycardia by Pacemaker ProgrammingPacing and Clinical Electrophysiology, 1982
- Pacemaker-induced ventricular tachycardia in patients with acute inferior myocardial infarctionInternational Journal of Cardiology, 1981
- Anodal stimulation as a cause of pacemaker-induced ventricular fibrillationAmerican Heart Journal, 1973
- VENTRICULAR FIBRILLATION AND PACINGAnnals of the New York Academy of Sciences, 1969
- PACING AND VENTRICULAR FIBRILLATIONAnnals of the New York Academy of Sciences, 1969
- Repetitive Ventricular Arrhythmia Resulting from Artificial Internal PacemakerCirculation, 1964
- VENTRICULAR FIBRILLATION DUE TO SINGLE, LOCALIZED INDUCTION AND CONDENSER SHOCKS APPLIED DURING THE VULNERABLE PHASE OF VENTRICULAR SYSTOLEAmerican Journal of Physiology-Legacy Content, 1940