Comparison of Two Myoepicardial Pacemaker Leads: Follow‐up in 80 Children, Adolescents, and Young Adults
Authors:
- 1 December 1988
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 11 (12) , 2216-2222
Abstract
Although severai types of commercially available epicardial leads exist, few postimplantation data have been reported. To compare “screw‐in” (6917–35) leads with “stab‐on” leads (4951–35) from the same manufacturer, we reviewed the records of 80 young patients (age 8 days to 29 years) who underwent ventricular epicardial pacemaker implantation from 1973 to 1986. Follow‐up for the 57 patients with the 6917–35 model ranged from 3 months to 17 years (median 6.5 years) and for the 23 patients with the 4951–35 model 9 days to 4.25 years (median 2.0 years). Actuarial life table analysis revealed significantly (P < 0.001) fewer 4951–35 leads were functioning at each of 1–5 years after implant, compared to the 6917–35 leads. Analysis of available threshold pulse width data revealed no difference (P = 0.08) acutely (6 weeks after implant), but a significantly (P = 0.05) higher mean threshold for the 4951–35 leads was found chronically. No significant correlation was found for lead failure with age, underlying heart disease, lead site (i.e., left or right ventricle), or surgical approach. Using the sutureless, stab‐on technique, the 4951–35 lead is associated with higher thresholds and lower survival rate when compared to the 6917–35 lead.Keywords
This publication has 15 references indexed in Scilit:
- Epicardial ventricular pacemaker electrode longevity in childrenThe American Journal of Cardiology, 1988
- Long-term follow-up of transvenous cardiac pacing in childrenThe American Journal of Cardiology, 1986
- Permanent Cardiac Pacing After Open Heart Surgery: Congenital Heart DiseasePacing and Clinical Electrophysiology, 1985
- Cardiac pacing in children: A 15-year experienceAmerican Heart Journal, 1985
- Experience with a New Epimyocardial Pacing Lead in ChildrenPacing and Clinical Electrophysiology, 1984
- One Turn More: Threshold Superiority of 3‐turn versus 2‐turn Screw‐in Myocardial ElectrodesPacing and Clinical Electrophysiology, 1984
- Transvenous pacing in pediatric patientsAmerican Heart Journal, 1983
- Experimental and Clinical Study of a New Permanent Myocardial Atrial Sutureless Pacing LeadPacing and Clinical Electrophysiology, 1983
- Ventricular Pacing in ChildrenPacing and Clinical Electrophysiology, 1982
- Fourteen Years of Implanted Pacemakers in ChildrenThe Annals of Thoracic Surgery, 1978