Keratometry with five different techniques: a study of device repeatability and inter-device agreement
- 23 February 2014
- journal article
- Published by Springer Nature in International Ophthalmology
- Vol. 34 (4) , 869-875
Abstract
The purpose of this study was to determine and compare the repeatability of keratometry measurements with the Javal manual keratometer, Topcon automated kerato-refractometer, IOLMaster, EyeSys Corneal Analysis System, and Pentacam Scheimpflug Topography System, and determine the agreement of measurements of the latter four devices with the Javal keratometer as the gold standard. In this cross-sectional study, 21 people with no history of ocular disease or surgery were examined twice with each device. Minimum, maximum, and average keratometry readings were recorded in diopters (D) for each eye and used in the analyses. For statistical analysis, we determined correlation coefficients and used the Bland–Altman method and calculated the 95 % limits of agreement (LoA). All repeatability coefficients were satisfactorily high. Best repeatability for minimum keratometry and maximum keratometry readings was seen with IOLMaster (95 % LoA −0.23 to 0.19 D and −0.31 to 0.32 D, respectively). Agreement with Javal manual keratometry was best with Topcon and IOLMaster for minimum keratometry readings (95 % LoA −0.67 to 0.28 D and −0.57 to 0.38 D, respectively), and with Topcon for maximum keratometry readings (95 % LoA −0.85 to 0.63 D). In our series of normal eyes, keratometry readings with Topcon, IOLMaster, and Pentacam showed very good agreement with Javal measurements, and inter-device agreements were better than the repeatability of the manual gold standard. However, differences between Javal and EyeSys may not be clinically acceptable, and these two devices should not be used interchangeably. Further studies are needed to investigate these issues in non-virgin eyes.Keywords
This publication has 40 references indexed in Scilit:
- Which Keratometer is Most Reliable for Correcting Astigmatism with Toric Intraocular Lenses?Korean Journal of Ophthalmology, 2012
- Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometryJournal of Cataract & Refractive Surgery, 2011
- Clinical evaluation of the Grand Seiko Auto Ref/Keratometer WAM-5500Ophthalmic and Physiological Optics, 2010
- Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive SurgeryJournal of Refractive Surgery, 2009
- Accuracy of a new partial coherence interferometry analyser for biometric measurementsBritish Journal of Ophthalmology, 2009
- Cataract Surgery in KeratoconusEye & Contact Lens: Science & Clinical Practice, 2007
- Calculation of intraocular lens power: a reviewActa Ophthalmologica Scandinavica, 2007
- Comparison of keratometric values of healthy eyes measured by javal keratometer, nidek autokeratometer, and corneal analysis system (EyeSys)International Contact Lens Clinic, 2000
- A critical discussion of intraclass correlation coefficientsStatistics in Medicine, 1994
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986