Evaluation of radiochromic gel dosimetry and polymer gel dosimetry in a clinical dose verification
- 22 August 2013
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 58 (18) , 6241-6262
Abstract
A quantitative comparison of two full three-dimensional (3D) gel dosimetry techniques was assessed in a clinical setting: radiochromic gel dosimetry with an in-house developed optical laser CT scanner and polymer gel dosimetry with magnetic resonance imaging (MRI). To benchmark both gel dosimeters, they were exposed to a 6 MV photon beam and the depth dose was compared against a diamond detector measurement that served as golden standard. Both gel dosimeters were found accurate within 4% accuracy. In the 3D dose matrix of the radiochromic gel, hotspot dose deviations up to 8% were observed which are attributed to the fabrication procedure. The polymer gel readout was shown to be sensitive to B0 field and B1 field non-uniformities as well as temperature variations during scanning. The performance of the two gel dosimeters was also evaluated for a brain tumour IMRT treatment. Both gel measured dose distributions were compared against treatment planning system predicted dose maps which were validated independently with ion chamber measurements and portal dosimetry. In the radiochromic gel measurement, two sources of deviations could be identified. Firstly, the dose in a cluster of voxels near the edge of the phantom deviated from the planned dose. Secondly, the presence of dose hotspots in the order of 10% related to inhomogeneities in the gel limit the clinical acceptance of this dosimetry technique. Based on the results of the micelle gel dosimeter prototype presented here, chemical optimization will be subject of future work. Polymer gel dosimetry is capable of measuring the absolute dose in the whole 3D volume within 5% accuracy. A temperature stabilization technique is incorporated to increase the accuracy during short measurements, however keeping the temperature stable during long measurement times in both calibration phantoms and the volumetric phantom is more challenging. The sensitivity of MRI readout to minimal temperature fluctuations is demonstrated which proves the need for adequate compensation strategies.Keywords
This publication has 15 references indexed in Scilit:
- Polymer gel dosimetryPhysics in Medicine & Biology, 2010
- Radiochromic leuco dye micelle hydrogels: I. Initial investigationPhysics in Medicine & Biology, 2009
- The influence of cooling rate on the accuracy of normoxic polymer gel dosimetersPhysics in Medicine & Biology, 2007
- The history and principles of optical computed tomography for scanning 3-D radiation dosimetersJournal of Physics: Conference Series, 2006
- Three dimensional radiation dosimetry in lung‐equivalent regions by use of a radiation sensitive gel foam: Proof of principleMedical Physics, 2006
- The fundamental radiation properties of normoxic polymer gel dosimeters: a comparison between a methacrylic acid based gel and acrylamide based gelsPhysics in Medicine & Biology, 2006
- Evaluation of the gamma dose distribution comparison methodMedical Physics, 2003
- Artefacts in multi-echoT2imaging for high-precision gel dosimetry: III. Effects of temperature drift during scanningPhysics in Medicine & Biology, 2001
- Validation of MR-based polymer gel dosimetry as a preclinical three-dimensional verification tool in conformal radiotherapyMagnetic Resonance in Medicine, 2000
- Mathematical analysis and experimental investigation of noise in quantitative magnetic resonance imaging applied in polymer gel dosimetrySignal Processing, 1998