How many times can we use a phosphor plate? A preliminary study
Authors:
- S ErgünSES Ergün
- D İlgüyDID İlgüy
- M İlgüyMIM İlgüy
- H BoyacıoğluHBH Boyacıoğlu
- 1 January 2009
- journal article
- research article
- Published by Oxford University Press (OUP) in Dentomaxillofacial Radiology
- Vol. 38 (1) , 42-47
Abstract
Objectives: Digital radiography has become a useful tool in daily dental practice due to the advances in imaging technologies. Charge coupled devices (CCDs) and photostimulable phosphor plates (PSPs) are currently in use for dental imaging; however, the longevity of PSPs in dental practice is not yet established. The aim of this study was to determine the service life of PSPs in a clinical setting. Methods: Five unused PSPs were exposed with a conventional X-ray device and converted into digital images with Digora Optime (Soredex, Milwaukee, WI). These were recorded as the baseline images. Subsequent digital images of the plates were obtained after 20, 40, 60, 80, 100, 120, 140, 160, 180 and 200 exposures. All radiographic images were subtracted from the first digital image (baseline) and the mean grey values (MGVs) of the subtracted images were established using software. The data were grouped in 3 classes according to the number of exposures (20-80; 100-140; 160-200), and were analysed using variance analysis and chi(2) tests. Results: The MGVs of the subtracted images varied between 126.25 and 127.59, and the difference was not significant among the groups (P = 0.11). However, the differences between the MGVs of the plates oil each exposure settings were significantly different than those of the baseline image (P < 0.05). Conclusions: The findings of this study revealed that even though a slight deterioration occurred after the first exposure, each plate can be used up to 200 times. Further studies are required to reach a more concrete conclusion.Keywords
This publication has 10 references indexed in Scilit:
- How does signal fade on photo-stimulable storage phosphor imaging plates when scanned with a delay and what is the effect on image quality?Oral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2006
- Clinical efficacy of a new software developed for dental digital subtraction radiographyDentomaxillofacial Radiology, 2006
- ROI-based image registration for digital subtraction radiographyOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2006
- Filmless imagingThe Journal of the American Dental Association, 2005
- Effect of delayed scanning of storage phosphor platesOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2005
- A digital subtraction radiography scheme based on automatic multiresolution registrationDentomaxillofacial Radiology, 2004
- Storage Phosphor Plates: How Durable are they as a Digital Dental Radiographic System?The Journal of Contemporary Dental Practice, 2004
- Analysis of digital images acquired using different phosphor storage plates (PSPs) subjected to varying reading times and storage conditionsDentomaxillofacial Radiology, 2003
- Digital Radiography: An OverviewThe Journal of Contemporary Dental Practice, 2002
- A new method for the automated alignment of dental radiographs for digital subtraction radiographyDentomaxillofacial Radiology, 2000