Radio frequency energy harvesting from a feeding source in a passive deep brain stimulation device for murine preclinical research
- 28 August 2015
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 37 (10) , 1020-1026
Abstract
This paper presents the development of an energy harvesting circuit for use with a head-mountable deep brain stimulation (DBS) device. It consists of a circular planar inverted-F antenna (PIFA) and a Schottky diode-based Cockcroft-Walton 4-voltage rectifier. The PIFA has the volume of π × 10(2) × 1.5 mm(3), resonance frequency of 915 MHz, and bandwidth of 16 MHz (909-925 MHz) at a return loss of -10 dB. The rectifier offers maximum efficiency of 78% for the input power of -5 dBm at a 5 kΩ load resistance. The developed rectenna operates efficiently at 915 MHz for the input power within -15 dBm to +5 dBm. For operating a DBS device, the DC voltage of 2 V is recorded from the rectenna terminal at a distance of 55 cm away from a 26.77 dBm transmitter in free space. An in-vitro test of the DBS device is presented.Keywords
This publication has 10 references indexed in Scilit:
- An inexpensive, charge-balanced rodent deep brain stimulation device: A step-by-step guide to its procurement and constructionJournal of Neuroscience Methods, 2013
- Long-Term High Frequency Deep Brain Stimulation of the Nucleus Accumbens Drives Time-Dependent Changes in Functional Connectivity in the Rodent Limbic SystemBrain Stimulation, 2013
- A Low Power Micro Deep Brain Stimulation Device for Murine Preclinical ResearchIEEE Journal of Translational Engineering in Health and Medicine, 2013
- Deep brain stimulation for treatment-resistant depression: Efficacy, safety and mechanisms of actionNeuroscience & Biobehavioral Reviews, 2012
- Wireless implantable micro-stimulation device for high frequency bilateral deep brain stimulation in freely moving miceJournal of Neuroscience Methods, 2012
- Strategy for Microwave Energy Harvesting From Ambient Field or a Feeding SourceIEEE Transactions on Power Electronics, 2012
- RF energy harvesting system and circuits for charging of mobile devicesIEEE Transactions on Consumer Electronics, 2010
- A fully implantable stimulator for use in small laboratory animalsJournal of Neuroscience Methods, 2007
- A stimulator with wireless power and signal transmission for implantation in animal experiments and other applicationsJournal of Neuroscience Methods, 1998
- Small-signal model of the Cockcroft-Walton voltage multiplierIEEE Transactions on Power Electronics, 1994