Continuous, Real-Time, Noninvasive Hemodynamic Cardiac Doppler Monitoring With a Novel Hands-Free Device: A Feasibility Study Compared to Standard Echo
- 27 June 2017
- journal article
- Published by ASME International in Journal of Medical Devices
- Vol. 11 (3) , 031001
Abstract
This study sought to determine the feasibility of using noninvasive cardiac hemodynamics (NICHE), a new noninvasive Doppler-based device, to monitor real-time, simultaneous tissue and blood-flow Doppler measurements in a clinical setting, and to obtain preliminary performance data compared to a commercially available system. Doppler-based measurements have been shown to correlate well with invasive hemodynamic data and diastolic function, but their use in clinical applications has been limited by various technical issues. The NICHE device was developed to obtain simultaneous tissue and blood-flow Doppler measurements automatically, in real-time and in a hands-free manner. Thirty participants (ten normal volunteers and 20 patients in a cardiac rehab program) underwent standard echocardiographic/Doppler studies followed immediately by NICHE monitoring. Early diastolic transmitral blood-flow velocity (E) and tissue Doppler myocardial wall velocity during early relaxation (E′) were acquired using a standard echo device; and E/E′ was derived post hoc. NICHE measurements included E, E′, and directly measured instantaneous E/E′. NICHE was successfully used in 28 participants. Measurements of ENICHE ranged from 40 cm/s to over 120 cm/s and correlated well with Eecho (R = 0.93). ranged from 2 to 23 cm/s and correlated well with the averaged (R = 0.91). Directly measured E/ ratios ranged from 3 to 23 and correlated well with derived E/ (R = 0.91). The NICHE device can monitor patients in a hands-free manner and can supply real-time Doppler derived measurements of hemodynamic parameters and diastolic function that correlate well with measurements from standard echo devices.
Keywords
This publication has 27 references indexed in Scilit:
- Noninvasive Cardiac Output Monitors: A State-of the-Art ReviewJournal of Cardiothoracic and Vascular Anesthesia, 2013
- Serial Doppler Echocardiography and Tissue Doppler Imaging in the Detection of Elevated Directly Measured Left Atrial Pressure in Ambulant Subjects With Chronic Heart FailureJACC: Cardiovascular Imaging, 2011
- Minimally Invasive Cardiac Output Monitoring in the Perioperative SettingAnesthesia & Analgesia, 2009
- Recommendations for the Evaluation of Left Ventricular Diastolic Function by EchocardiographyJournal of the American Society of Echocardiography, 2009
- Doppler tissue imaging for the assessment of left ventricular diastolic function: A systematic approach for the sonographerJournal of the American Society of Echocardiography, 2005
- Minimally invasive hemodynamic monitoring for the intensivist: Current and emerging technologyCritical Care Medicine, 2002
- Clinical Utility of Doppler Echocardiography and Tissue Doppler Imaging in the Estimation of Left Ventricular Filling PressuresCirculation, 2000
- Doppler Estimation of Left Ventricular Filling Pressure in Sinus TachycardiaCirculation, 1998
- Doppler Tissue Imaging Quantitates Regional Wall Motion During Myocardial Ischemia and ReperfusionCirculation, 1998
- An Index of Early Left Ventricular Filling That Combined With Pulsed Doppler Peak E Velocity May Estimate Capillary Wedge PressureJACC, 1997