Rebound Tonometer Compared With Goldmann Tonometer in Normal and Pathologic Corneas
Authors:
- Javier Moreno-MontañésJMJavier Moreno-Montañés
- Noelia GarcíaNGNoelia García
- Ana Fernández-HortelanoAFAna Fernández-Hortelano
- Alfredo García-LayanaAGAlfredo García-Layana
- 1 May 2007
- journal article
- research article
- Published by Wolters Kluwer Health in Cornea
- Vol. 26 (4) , 427-430
Abstract
To compare the intraocular pressure (IOP) measurements obtained with the rebound tonometer (RT) and the Goldmann handheld tonometer (GT) in normal and altered corneas. A total of 208 normal corneas and 50 corneas with pathologies were included in this prospective study. All measurements were randomly obtained by 1 observer. The medians and interquartile range (IR) for both tonometers were compared. The median differences were assessed in IOP groups. Agreement between the tonometers was calculated using the Bland-Altman method. The median IOP in all eyes was 17 mm Hg (IR, 13-22 mm Hg) with the RT and 16 mm Hg (IR, 13-21 mm Hg) with the GT (P < 0.001). The correlation was excellent between tonometers (r2 = 0.86; P < 0.001). The minimal differences between the two were obtained from 10 to 20 mm Hg (GT). The Bland-Altman scatterplot obtained good agreement between the instruments. In normal corneas, the median difference was < or =2 mm Hg in 77.4% of cases. In the altered corneas, the median difference was < or =2 mm Hg in 73% of cases (P = 0.21 compared with the normal group). In 10% and 2% of cases, the IOP could not be measured using the GT and RT, respectively. The results were similar for both tonometers. In the altered corneas, the IOP could be difficult to obtain with the GT because of distorted half-circles. The 1-mm-diameter disposable RT tip facilitated obtaining measurements without anesthetic drops, which avoids infections. The RT could be useful in routine clinical settings when measuring IOP in corneas with pathologies.Keywords
This publication has 22 references indexed in Scilit:
- Noninvasive Measurement of Rodent Intraocular Pressure with a Rebound TonometerInvestigative Ophthalmology & Visual Science, 2005
- Central corneal thickness and measured IOP response to topical ocular hypotensive medication in the Ocular Hypertension Treatment StudyAmerican Journal of Ophthalmology, 2004
- Method for the Noninvasive Measurement of Intraocular Pressure in MiceInvestigative Ophthalmology & Visual Science, 2003
- Influence of Corneal Variables on Accuracy of Intraocular Pressure MeasurementJournal of Glaucoma, 2003
- Central Corneal Thickness in Normal, Glaucomatous, and Ocular Hypertensive EyesArchives of Ophthalmology (1950), 1997
- REGRET GRAPHS, DIAGNOSTIC UNCERTAINTY AND YOUDEN'S INDEXStatistics in Medicine, 1996
- Number of people with glaucoma worldwide.British Journal of Ophthalmology, 1996
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- OCULAR RIGIDITY AND INTRAOCULAR PRESSURE IN KERATOCONUSAustralian Journal of Opthalmology, 1984
- Über ApplanationstonometrieOphthalmologica, 1957