Bioprosthetic Valve Remodeling of Trifecta Surgical Valves to Facilitate Valve-in-Valve TAVR
Authors:
- John T. SaxonJSJohn T. Saxon
- David J. CohenDCDavid J. Cohen
Open Access
- 1 March 2020
- journal article
- research article
- Published by Elsevier in Structural Heart
- Vol. 4 (2) , 99-104
Abstract
Background: Valve-in-valve transcatheter aortic valve replacement (VIV TAVR) is an effective treatment for high-risk patients with failed surgical bioprosthetic valves, but may be complicated by patient-prosthesis mismatch (PPM). We describe a novel technique – bioprosthetic valve remodeling (BVR) – to distort the frame of Trifecta surgical valves with a high-pressure balloon inflation, allowing for improved expansion and hemodynamics of the transcatheter heart valve (THV). Methods: We performed bench testing of BVR by deploying Portico THVs (Abbott, Minneapolis, MN, USA) inside Trifecta GT valves both at baseline and following BVR. Fluoroscopic and video imaging was performed to assess Trifecta frame distortion, THV expansion, and leaflet motion. Hydrodynamic testing was performed to determine transvalvular gradient under baseline conditions and following BVR. We performed 6 clinical cases of VIV TAVR with Evolut R THVs (Medtronic, Minneapolis, MN), followed by BVR. Hemodynamic measurements and calculation of the effective orifice area (EOA) were performed at baseline, immediately after implantation of the THV, and after BVR. Results: BVR resulted in an increased in the post-to-post diameter in Trifecta valves, and a lower transvalvular gradient after THV implantation. VIV TAVR and BVR was successful in all cases, with no major short-term complications. The mean baseline transvalvular gradient was 47.5 ± 14.4 mmHg, which was reduced to 21.0 ± 6.2 mmHg after THV implantation, and further reduced to 11.7 ± 4.1 mmHg following BVR. Gradients were stable at 1 month follow up. Conclusion: BVR distorts the frame of Trifecta valves, improving THV expansion and procedural hemodynamics of VIV TAVR.Keywords
This publication has 18 references indexed in Scilit:
- Transcatheter Aortic Valve Implantation Within Degenerated Aortic Surgical BioprosthesesJACC, 2017
- In vitro evaluation of implantation depth in valve-in-valve using different transcatheter heart valvesEuroIntervention, 2016
- Predictors and Outcomes of Prosthesis-Patient Mismatch After Aortic Valve ReplacementJACC: Cardiovascular Imaging, 2016
- Prosthesis-Patient MismatchJACC: Cardiovascular Imaging, 2016
- When valve-in-valve implantation is not sufficient: Bioprosthetic Russian dollsThe Journal of Thoracic and Cardiovascular Surgery, 2016
- Fracturing the Ring of Small Mitroflow Bioprostheses by High-Pressure Balloon Predilatation in Transcatheter Aortic Valve-in-Valve ImplantationCirculation: Cardiovascular Interventions, 2015
- Cracking the ring of Edwards Perimount bioprosthesis with ultrahigh pressure balloons prior to transcatheter valve in valve implantationInternational Journal of Cardiology, 2014
- Transcatheter Aortic Valve Implantation in Failed Bioprosthetic Surgical ValvesJAMA, 2014
- Transcatheter Aortic Valve-in-Valve Implantation for Patients With Degenerative Surgical Bioprosthetic ValvesCurrent Problems in Cardiology, 2014
- Prosthesis-patient mismatch: definition, clinical impact, and preventionHeart, 2005