23-gauge versus 20-gauge system for pars plana vitrectomy: a prospective randomised clinical trial

Authors:

Abstract
Aim: To compare the sutureless 23-gauge system with a standard 20-gauge system in pars plana vitrectomy. Methods: 60 patients in two randomised groups were included in this prospective clinical trial. Pars plana vitrectomy with either 23- or 20-gauge instruments was performed. The main outcome measures were postoperative conjunctival injection and pain. Secondary outcome parameters were time of surgery, intraocular pressure, visual acuity and complications. Results: Conjunctival injection (p = 0.0003) and postoperative pain (p = 0.01) were significantly reduced following 23-gauge vitrectomy compared with the 20-gauge procedure. Opening (p = 0.006) and closure times (pConclusion: The 23-gauge system for pars plana vitrectomy offers significantly higher patient comfort during the early postoperative period. Time of surgery is almost equal—a shorter time for wound closure is neutralised by a longer vitrectomy time in the 23-gauge group.