Magnetization transfer effect on the creatine methyl resonance studied by CW off‐resonance irradiation in human skeletal muscle on a clinical MR system

Authors:

Abstract
Magnetization transfer (MT) between the mobile (MR‐visible) spin pool and immobile (MR‐invisible) spin pool of creatine (Cr) was studied on a clinical 1.5 T MR scanner in human skeletal muscle using continuous wave (CW) pre‐irradiation as the saturation method for the immobile pool. For this purpose, only slight modifications to the MR system were made. A specially designed electronic circuit was used to couple a CW amplifier to the RF channel of the scanner. The CW pulse power (γB2/2π) and pulse length were determined to be approximately 550 Hz and 3 s, respectively, for optimal signal attenuation of the Cr methyl signal. The bound Cr fraction in human gastrocnemius muscle was determined to be 0.4–1.3% using a two‐pool exchange model function to describe the MT effect. Magn Reson Med 50:468–473, 2003.