Impact of temporal resolution on perfusion metrics, therapy decision, and radiation dose reduction in brain CT perfusion in patients with suspected stroke
Authors:
Open Access
- 18 March 2024
- journal article
- research article
- Published by Springer Nature in Neuroradiology
- Vol. 66 (5) , 749-759
Abstract
Purpose: CT perfusion of the brain is a powerful tool in stroke imaging, though the radiation dose is rather high. Several strategies for dose reduction have been proposed, including increasing the intervals between the dynamic scans. We determined the impact of temporal resolution on perfusion metrics, therapy decision, and radiation dose reduction in brain CT perfusion from a large dataset of patients with suspected stroke. Methods: We retrospectively included 3555 perfusion scans from our clinical routine dataset. All cases were processed using the perfusion software VEOcore with a standard sampling of 1.5 s, as well as simulated reduced temporal resolution of 3.0, 4.5, and 6.0 s by leaving out respective time points. The resulting perfusion maps and calculated volumes of infarct core and mismatch were compared quantitatively. Finally, hypothetical decisions for mechanical thrombectomy following the DEFUSE-3 criteria were compared. Results: The agreement between calculated volumes for core (ICC = 0.99, 0.99, and 0.98) and hypoperfusion (ICC = 0.99, 0.99, and 0.97) was excellent for all temporal sampling schemes. Of the 1226 cases with vascular occlusion, 14 (1%) for 3.0 s sampling, 23 (2%) for 4.5 s sampling, and 63 (5%) for 6.0 s sampling would have been treated differently if the DEFUSE-3 criteria had been applied. Reduction of temporal resolution to 3.0 s, 4.5 s, and 6.0 s reduced the radiation dose by a factor of 2, 3, or 4. Conclusion: Reducing the temporal sampling of brain perfusion CT has only a minor impact on image quality and treatment decision, but significantly reduces the radiation dose to that of standard non-contrast CT.Keywords
Funding Information
- Universitätsklinikum Freiburg
This publication has 38 references indexed in Scilit:
- CT perfusion with increased temporal sampling interval to predict target mismatch status in patients with acute ischemic strokeNeuroradiology, 2022
- Machine Outputs Must Be CheckedClinical Neuroradiology, 2021
- Automated Perfusion Calculations vs. Visual Scoring of Collaterals and CBV-ASPECTSClinical Neuroradiology, 2020
- Radiation exposure of computed tomography imaging for the assessment of acute strokeNeuroradiology, 2020
- Comparison of Automated CT Perfusion Softwares in Evaluation of Acute Ischemic StrokeJournal of Stroke and Cerebrovascular Diseases, 2019
- CT perfusion in acute stroke: Know the mimics, potential pitfalls, artifacts, and technical errorsEmergency Radiology, 2013
- Real‐time diffusion‐perfusion mismatch analysis in acute strokeJournal of Magnetic Resonance Imaging, 2010
- Increasing Sampling Interval in Cerebral Perfusion CTAcademic Radiology, 2010
- Dose reduction in dynamic perfusion CT of the brain: effects of the scan frequency on measurements of cerebral blood flow, cerebral blood volume, and mean transit timeEuropean Radiology, 2008
- Dynamic perfusion CT: Optimizing the temporal resolution for the calculation of perfusion CT parameters in stroke patientsEuropean Journal of Radiology, 2007