Analysis of aortic valve commissural fusion after support with continuous-flow left ventricular assist device
Authors:
- 24 June 2013
- journal article
- Published by Oxford University Press (OUP) in Interactive CardioVascular and Thoracic Surgery
- Vol. 17 (4) , 616-624
Abstract
Continuous-flow left ventricular assist devices (cf-LVADs) may induce commissural fusion of the aortic valve leaflets. Factors associated with this occurrence of commissural fusion are unknown. The aim of this study was to examine histological characteristics of cf-LVAD-induced commissural fusion in relation to clinical variables. Gross and histopathological examinations were performed on 19 hearts from patients supported by either HeartMate II (n = 17) or HeartWare (n = 2) cf-LVADs and related to clinical characteristics (14 heart transplantation, 5 autopsy). Eleven of the 19 (58%) aortic valves showed fusion of single or multiple commissures (total fusion length 11 mm [4-20] (median [interquartile range]) per valve), some leading to noticeable nodular displacements or considerable lumen diameter narrowing. Multiple fenestrations were observed in one valve. Histopathological examination confirmed commissural fusion, with varying changes in valve layer structure without evidence of inflammatory infiltration at the site of fusion. Commissural fusion was associated with continuous aortic valve closure during cf-LVAD support (P = 0.03). LVAD-induced aortic valve insufficiency developed in all patients with commissural fusion and in 67% of patients without fusion. Age, duration of cf-LVAD support and aetiology of heart failure (ischaemic vs dilated cardiomyopathy) were not associated with the degree of fusion. Aortic valve commissural fusion after support with cf-LVADs is a non-inflammatory process leading to changes in valve layer structure that can be observed in >50% of cf-LVAD patients. This is the first study showing that patients receiving full cf-LVAD support without opening of the valve have a significantly higher risk of developing commissural fusion than patients on partial support.Keywords
This publication has 20 references indexed in Scilit:
- Complete Aortic Valve Fusion After HeartMate II Left Ventricular Assist Device SupportJournal of Cardiothoracic and Vascular Anesthesia, 2012
- Clinical Experience With HeartWare Left Ventricular Assist Device in Patients With End-Stage Heart FailureThe Annals of Thoracic Surgery, 2012
- Destination therapy with left ventricular assist devices: patient selection and outcomesCurrent Opinion in Cardiology, 2011
- Geometry and Fusion of Aortic Valves from Pulsatile Flow Ventricular Assist Device Patients2011
- Reversal of Severe Heart Failure With a Continuous-Flow Left Ventricular Assist Device and Pharmacological TherapyCirculation, 2011
- The Development of Aortic Insufficiency in Left Ventricular Assist Device-Supported PatientsCirculation: Heart Failure, 2010
- Hemodynamic Stress Echocardiography in Patients Supported With a Continuous-Flow Left Ventricular Assist DeviceJACC: Cardiovascular Imaging, 2010
- Biomechanics of the Aortic Valve in the Continuous Flow VAD-Assisted HeartASAIO Journal, 2010
- Extended Mechanical Circulatory Support With a Continuous-Flow Rotary Left Ventricular Assist DeviceJACC, 2009
- Fusion of Aortic Valve Commissures in Patients Supported by a Continuous Axial Flow Left Ventricular Assist DeviceThe Journal of Heart and Lung Transplantation, 2008