Automatic patient centering in computed tomography: a systematic review and meta-analysis
Authors:
- 21 November 2024
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 35 (6) , 3486-3498
Abstract
Objective To comprehensively examine the influence of auto-patient centering technologies on positioning accuracy, radiation dose, image quality, and time efficiency of computed tomography (CT) scans. Materials and methods A systematic search of peer-reviewed English publications was performed between January 2000 and November 2023 in PubMed, Embase, CINAHL, Scopus, and Web of Science. Two postgraduate students and an academic lecturer independently reviewed the articles to verify adherence to the inclusion criteria. The QUADAS-2 tool was employed to evaluate study quality. We derived summary estimates on positioning accuracy, radiation dose reduction, image quality, and time efficiency using proportion and meta-analysis methodologies. Results Nine studies were identified comparing automatic and manual CT positioning. Automatic positioning improved accuracy by reducing vertical offsets to 7 mm and 4 mm for thorax and abdominal CTs, compared to 19 mm and 18 mm with manual methods. Most studies showed significant reductions in radiation dose, ranging from 5.71 to 31%. Image quality results were mixed, automatic methods generally produced images with less noise, but differences were minimal. Time efficiency was better, with automatic positioning reducing preparation time from 0.48 min versus 0.67 min for manual positioning. Conclusions This review confirms that automatic patient-centering technologies enhance positioning accuracy and decrease preparation times in CT scans. While reductions in radiation doses and some improvements in image quality were observed, the evidence remains mixed. Findings support integrating these technologies into clinical practice to optimize patient care. Key Points Question Does automatic patient centering in CT enhance positioning accuracy, reduce radiation exposure, and improve image quality? Findings Findings indicate that automatic centering can optimize image quality, reduce examination times and contribute to overall improvements in imaging efficiency. Clinical relevance Automatic patient centering in CT improves positioning accuracy, minimizes radiation exposure, enhances image quality, and accelerates imaging workflows, contributing to safer, more efficient imaging procedures that benefit patient care. Graphical AbstractKeywords
This publication has 40 references indexed in Scilit:
- Investigating the CT localizer radiograph: acquisition parameters, patient centring and their combined influence on radiation doseBritish Journal of Radiology, 2015
- Effect of Patient Centering on Patient Dose and Image Noise in Chest CTAmerican Journal of Roentgenology, 2014
- Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT—phantom studyPediatric Radiology, 2013
- Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studiesPhysica Medica, 2012
- ALARA (as low as reasonably achievable) CT 2011—executive summaryPediatric Radiology, 2011
- How to write a Cochrane systematic reviewNephrology, 2010
- Efficient use of automatic exposure control systems in computed tomography requires correct patient positioningActa Radiologica, 2009
- Application of Automatic Vertical Positioning Software to Reduce Radiation Exposure in Multidetector Row Computed Tomography of the ChestInvestigative Radiology, 2008
- Radiation Risk to Children From Computed TomographyPediatrics, 2007
- The influence of patient centering on CT dose and image noiseMedical Physics, 2007