Simulator-assisted Setting of Scan Protocols for X-ray CT: Development and Clinical Usefulness of the Scan Plan Simulator
Authors:
Open Access
- 1 January 2006
- journal article
- Published by Japanese Society of Radiological Technology in Japanese Journal of Radiological Technology
- Vol. 62 (1) , 95-104
Abstract
To assist in the selection of complicated computed tomography (CT) scan protocols and to obtain stable image SD values, prototype “Scan Plan Simulator” software with the following functions was developed and evaluated. 1) The image SD value that will be obtained in actual scanning is estimated by entering the patient's body size and scan protocol, after which a simulated image is displayed so that the estimated image SD value can be checked. The exposure dose can also be estimated in the Scan Plan Simulator. 2) The appropriate tube current is automatically set by entering the required image SD value. We evaluated the accuracy of the Scan Plan Simulator by comparing the simulation results with the image SD values and exposure dose obtained in actual scanning and assessed the usefulness of this method using Monte Carlo simulation based on body thickness data obtained in clinical examinations. The results showed that the Scan Plan Simulator not only stabilizes image SD by minimizing the effects of body size but also permits the exposure dose to be reduced by optimizing tube current.Keywords
This publication has 11 references indexed in Scilit:
- A method of predicting the image noise in paediatric multi-slice computed tomography imagesRadiation Protection Dosimetry, 2005
- Assessment of image quality of a standard and three dose-reducing protocols in adult cranial CTEuropean Radiology, 2004
- Radiation Dose and Image Quality in Pediatric CT: Effect of Technical Factors and Phantom Size and ShapeRadiology, 2004
- Reduced Radiation Dose Helical Chest CT: Effect on Reader Evaluation of Structures and Lung FindingsRadiology, 2004
- Low-Dose Chest CT: Optimizing Radiation Protection for PatientsAmerican Journal of Roentgenology, 2004
- Individually weight-adapted examination protocol in retrospectively ECG-gated MSCT of the heartEuropean Radiology, 2003
- Detection of Coronary Calcifications: Feasibility of Dose Reduction with a Body Weight–Adapted Examination ProtocolAmerican Journal of Roentgenology, 2003
- Computer-Simulated Radiation Dose Reduction for Abdominal Multidetector CT of Pediatric PatientsAmerican Journal of Roentgenology, 2002
- Individually Adapted Examination Protocols for Reduction of Radiation Exposure in Chest CTInvestigative Radiology, 2001
- MAPD—an objective way to select mAs for paediatric brain CTBritish Journal of Radiology, 2001