An Adaptable Computed Tomography-Derived Three-Dimensional-Printed Alignment Fixture Minimizes Errors in Radius Biomechanical Testing
Open Access
- 19 July 2021
- journal article
- research article
- Published by Wiley in Journal of Biomechanical Engineering
- Vol. 143 (11)
Abstract
Biomechanical testing of long bones can be susceptible to errors and uncertainty due to malalignment of specimens with respect to the mechanical axis of the test frame. To solve this problem, we designed a novel, customizable alignment and potting fixture for long bone testing. The fixture consists of three-dimensional-printed components modeled from specimen-specific computed tomography (CT) scans to achieve a predetermined specimen alignment. We demonstrated the functionality of this fixture by comparing benchtop torsional test results to specimen-matched finite element models and found a strong correlation (R2 = 0.95, p < 0.001). Additional computational models were used to estimate the impact of malalignment on mechanical behavior in both torsion and axial compression. Results confirmed that torsion testing is relatively robust to alignment artifacts, with absolute percent errors less than 8% in all malalignment scenarios. In contrast, axial testing was highly sensitive to setup errors, experiencing absolute percent errors up to 50% with off-center malalignment and up to 170% with angular malalignment. This suggests that whenever appropriate, torsion tests should be used preferentially as a summary mechanical measure. When more challenging modes of loading are required, pretest clinical-resolution CT scanning can be effectively used to create potting fixtures that allow for precise preplanned specimen alignment. This may be particularly important for more sensitive biomechanical tests (e.g., axial compressive tests) that may be needed for industrial applications, such as orthopedic implant design.Keywords
Funding Information
- National Science Foundation (CMMI-1943287)
This publication has 31 references indexed in Scilit:
- Reproducible reference frame for in vitro testing of the human vertebraeJournal of Biomechanics, 2014
- Evaluation of a Customized Artificial Osteoporotic Bone Model of the Distal FemurJournal of Biomaterials Applications, 2010
- Comparison of three standard anatomical reference frames for the tibia–fibula complexJournal of Biomechanics, 2008
- Mathematical relationships between bone density and mechanical properties: A literature reviewClinical Biomechanics, 2008
- Dependence of mechanical compressive strength on local variations in microarchitecture in cancellous bone of proximal human femurJournal of Biomechanics, 2008
- Intra-Articular Pressure Measurement in the Radioulnocarpal Joint Using a Novel Sensor: In Vitro and In Vivo ResultsThe Journal of Hand Surgery, 2007
- Mechanical testing of cancellous bone from the femoral head: Experimental errors due to off-axis measurementsJournal of Biomechanics, 2007
- A surrogate long-bone model with osteoporotic material properties for biomechanical testing of fracture implantsJournal of Biomechanics, 2007
- Trabecular bone modulus–density relationships depend on anatomic siteJournal of Biomechanics, 2003
- Torsional Strength of the Radius After Osteofasciocutaneous Free Flap Harvest with and Without Primary Bone PlatingOtolaryngology–Head and Neck Surgery, 2000