Biaxial flex‐fatigue and viral penetration of natural rubber latex gloves before and after artificial aging
- 1 January 2002
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 63 (6) , 739-745
Abstract
Barrier integrity of unaged and oven‐aged (at 70 °C) natural rubber latex examination gloves was assessed with a biaxial flex‐fatigue method where failure was detected electronically, and by live viral penetration testing performed according to a modified version of ASTM F1671‐97a. When no change in barrier properties was detected during flex testing, no virus passage was found after viral challenge. Conversely, when a change in the barrier properties was indicated by the electrical signal, virus passage was found in 74% of the specimens. Flex‐fatigue results indicated that unaged test specimens from powdered (PD) and powder‐free (PF) nonchlorinated gloves had significantly longer fatigue lives than powder‐free chlorinated (CL) gloves from the same manufacturer. Biaxial flexing of oven‐aged glove specimens showed a marginal increase in fatigue life for the PF gloves, but no increase for the PD gloves. The fatigue life of the CL gloves was observed to increase significantly after oven aging. However, this appears to be due to a design feature of the test apparatus, wherein peak volume displacement of the worked specimen is held constant. An aging‐induced change in the viscoelastic properties of the CL gloves—permanent deformation of the specimens early in the fatigue test—relieves the stress magnitude applied as the test progresses. Thus, permanent deformation acts as a confounding factor in measuring durability of latex gloves by fixed displacement flex‐fatigue. © 2002 Wiley Periodicals, Inc. J Biomed Mater Res (Appl Biomater) 63: 739–745, 2002Keywords
This publication has 9 references indexed in Scilit:
- An experimental technique for determining biaxial fatigue lifetimes in biomedical elastomersInternational Journal of Fatigue, 2001
- Efficacy of protection by latex gloves during orthodontic therapyJournal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie, 2000
- Modification of ASTM F 1671-97a, Resistance of Materials to Penetration by Blood-Borne Pathogens, for Use with Elastomeric MaterialsJournal of Testing and Evaluation, 2000
- Environmental Degradation of Natural Rubber Latex Gloves: The Effects of Elevated Temperature on Tensile StrengthJournal of Testing and Evaluation, 1999
- Ability of a Viral Penetration Test (ASTM F1671-95) to Detect Small HolesJournal of Testing and Evaluation, 1999
- Glove perforation rate in open lung surgeryEuropean Journal of Cardio-Thoracic Surgery, 1999
- Permeability and material characteristics of vulcanized latex film during and following cyclic fatigue in a saline environmentJournal of Applied Polymer Science, 1997
- The Effect of Glove Flexure on Permeation ParametersApplied Occupational and Environmental Hygiene, 1997
- Permeation of Cancer Chemotherapeutic Drugs Through Glove Materials under Static and Flexed ConditionsApplied Occupational and Environmental Hygiene, 1990