Automated detection of transient ST-segment episodes in 24h electrocardiograms
Authors:
- 1 May 2004
- journal article
- research article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 42 (3) , 303-311
Abstract
A novel automated system is presented for improved detection of transient ischaemic and heart rate-related ST-segment episodes in ‘real-world’ 24 h ambulatory ECG data. Using a combination of traditional time-domain and Karhunen-Loève transform-based approaches, the detector derives QRS complex and ST-segment morphology feature vectors and, by mimicking human examination of feature-vector time series and their trends, tracks the time-varying ST-segment reference level owing to clinically unimportant, non-ischaemic causes, such as slow drifts, axis shifts and conduction changes. The detector estimates the slowly varying ST-segment level trend, identifies step changes in the time series and subtracts the ST-segment reference level thus obtained from the ST-segment level to obtain the ST-segment deviation time series, which are suitable for detection of ST-segment episodes. The detector was developed using the Long-term ST database containing 24h ambulatory ECG records with human-expert annotated transient ischaemic and heart rate-related ST-segment episodes. The average ST episode detection sensitivity/positive predictivity obtained when using the annotations of the annotation protocol B of the database were 78.9%/80.7%. Evaluation of the detector using the European Society of Cardiology ST-T database as a test database showed average ST episode detection sensitivity/positive predictivity of 81.3%/89.2%, which are better performances, comparable with those of the systems being developed using the European database.Keywords
This publication has 19 references indexed in Scilit:
- Long-term ST database: A reference for the development and evaluation of automated ischaemia detectors and for the study of the dynamics of myocardial ischaemiaMedical & Biological Engineering & Computing, 2003
- Performance measures for algorithms to detect transient ischemic ST segment changes2002
- Hierarchical state space partitioning with a network self-organising map for the recognition of ST-T segment changesMedical & Biological Engineering & Computing, 2000
- Automatic detection of ST-T complex changes on the ECG using filtered RMS difference series: application to ambulatory ischemia monitoringIEEE Transactions on Biomedical Engineering, 2000
- Detection of Transient ST Segment Episodes During Ambulatory ECG MonitoringComputers and Biomedical Research, 1998
- Guidelines for assessing performance of ST analysers.Journal of Medical Engineering & Technology, 1998
- Developments in ECG acquisition, preprocessing, parameter measurement, and recordingIEEE Engineering in Medicine and Biology Magazine, 1998
- An approach to intelligent ischaemia monitoringMedical & Biological Engineering & Computing, 1995
- Computer system for analysis of ST segment changes on 24 hour Holter monitor tapes: Comparison with other available systemsJACC, 1984
- Bootstrap Methods: Another Look at the JackknifeThe Annals of Statistics, 1979