Unusual Case of AneuRx Stent-Graft Failure Two Years after AAA Exclusion
- 1 October 2001
- journal article
- case report
- Published by SAGE Publications in Journal of Endovascular Therapy
- Vol. 8 (5) , 465-471
Abstract
Purpose: To present a case of type-III endoleak due to separation of modular components and suture breakage after abdominal aortic aneurysm (AAA) repair with the AneuRx stent-graft. Case Report: A 58-year-old man with a 78-mm AAA underwent successful aneurysm exclusion with an AneuRx stent-graft in 1998. Because the bifurcated component migrated during deployment, an additional aortic cuff was deployed to extend the stent-graft proximally to just below the renal arteries. On routine evaluation with computed tomography at 2 years postimplantation, a type-III endoleak with anterior displacement of the stent-graft was discovered. Angiography revealed separation of the aortic cuff from the stent-graft body. The aneurysm, which had decreased to 70 mm at 6 months, had enlarged to 80 mm in diameter. The patient underwent surgical AAA repair with removal of the stent-graft. Macroscopic examination showed suture breakage and separation of the stent rings and the graft material. Conclusions: This late complication of the AneuRx endograft underscores the need for frequent imaging surveillance in all patients undergoing endoluminal AAA repair.Keywords
This publication has 14 references indexed in Scilit:
- Incidence and risk factors of late rupture, conversion, and death after endovascular repair of infrarenal aortic aneurysms: The EUROSTAR experienceJournal of Vascular Surgery, 2000
- Endoleak as a predictor of outcome after endovascular aneurysm repair: AneuRx multicenter clinical trialJournal of Vascular Surgery, 2000
- Aneurysm rupture after endovascular repair using the AneuRx stent graftJournal of Vascular Surgery, 2000
- Longitudinal Aneurysm Shrinkage following Endovascular Aortic Aneurysm Repair: A Source of Intermediate and Late ComplicationsJournal of Endovascular Therapy, 1999
- Symposium on Distortion and Structural Deterioration of Endovascular Grafts Used to Repair Abdominal Aortic AneurysmsJournal of Endovascular Therapy, 1999
- Endoleakage after stent-graft treatment of abdominal aneurysm: Implications on pressure and imaging—an in vitro studyJournal of Vascular Surgery, 1998
- Rupture of an abdominal aortic aneurysm after endovascular graft placement and aneurysm size reductionJournal of Vascular Surgery, 1998
- Rupture of abdominal aortic aneurysm previously treated by endovascular stentgraftJournal of Vascular Surgery, 1998
- Clinical consequences of periprosthetic leak after endovascular repair of abdominal aortic aneurysmJournal of Vascular Surgery, 1998
- A Prospective Study of Changes in Morphology and Dimensions of Abdominal Aortic Aneurysms following Endoluminal Repair: A Preliminary ReportJournal of Endovascular Therapy, 1995