Postoperative Intracochlear Electrocochleography in Pediatric Cochlear Implant Recipients: Association to Audiometric Thresholds and Auditory Performance
Authors:
- 21 January 2020
- journal article
- research article
- Published by Wolters Kluwer Health in Ear & Hearing
- Vol. 41 (5) , 1135-1143
Abstract
Objectives: The aim of this study was to compare intracochlear-recorded cochlear microphonics (CM) responses to behavioral audiometry thresholds in young children, with congenital hearing loss, 2 to 5 years after cochlear implantation early in life. In addition, differences in speech and auditory outcomes were assessed among children with and without residual hearing. Design: The study was conducted at a tertiary, university-affiliated, pediatric medical center. CM responses by intracochlear electrocochleography technique were recorded from 102 implanted ears of 60 children and those responses correlated to behavioral audiometry thresholds at 0.125 to 2 kHz frequencies. All children had received Advanced Bionics cochlear implant with High Focus J1 or MidScala electrodes, along with extensive auditory rehabilitation before and after implantation, including the use of conventional hearing aids. Speech Spatial and Hearing Qualities, Category of Auditory Performance scale, and educational settings information were obtained for each participant. Those cochlear implantation (CI) outcomes were compared between children with or without residual CM responses. Results: Two distinctive CM responses patterns were found among the implanted children. Of all ears diagnosed with cochlear hearing loss (n = 88), only in 29 ears, clear CM responses were obtained. In all other ears, no CM responses were obtained at the maximum output levels. The CM responses were highly correlated with coefficients ranging from 0.7 to 0.83 for the audiometric behavioral thresholds at 0.125 to 2 kHz frequency range. Of all ears diagnosed with auditory neuropathy syndrome disorder (n = 14), eight ears had residual hearing and recordable CM postimplantation. The other six ears showed no recordable CM responses at maximum output levels for all tested frequencies. Those showed supposedly better CM responses than the behavioral audiometry threshold, however with poor correlations with tested frequency thresholds. Children with residual hearing showed significantly better auditory outcomes with CI, compared with those without residual hearing. Conclusions: In children with congenital cochlear hearing loss, the objective CM intracochlear responses can reliably predict the residual audiometric threshold. However, in children with auditory neuropathy syndrome disorder, the CM thresholds did not match the behavioral audiometric responses. Postoperatively, children with recordable CM responses, showing preserved residual hearing, demonstrated better CI outcomes.Keywords
This publication has 46 references indexed in Scilit:
- Intraoperative Monitoring Using Cochlear Microphonics in Cochlear Implant Patients With Residual HearingOtology & Neurotology, 2012
- Resolution of Untoward Nonauditory Stimulation and Cochlear Implant Performance Gain After Extraction of a Stainless Steel Dental Crown in a Patient With Cochlear OtosclerosisOtology & Neurotology, 2011
- New Criteria of Indication and Selection of Patients to Cochlear ImplantInternational Journal of Otolaryngology, 2011
- Electrocochleography: A Review of Recording Approaches, Clinical Applications, and New Findings in Adults and ChildrenJournal of the American Academy of Audiology, 2010
- Multi-site diagnosis and management of 260 patients with Auditory Neuropathy/Dys-synchrony (Auditory Neuropathy Spectrum Disorder*)International Journal of Audiology, 2010
- Cochlear microphonic audiometry: a new hearing test for objective diagnosis of deafnessActa Oto-Laryngologica, 2009
- Electrical Excitation of the Acoustically Sensitive Auditory Nerve: Single-Fiber Responses to Electric Pulse TrainsJournal of the Association for Research in Otolaryngology, 2006
- Intracochlear Electrical Stimulation of Normal and DEAF CATS Investigated Using Brainstem Response AudiometryActa Oto-Laryngologica, 1983
- Diagnostic Significance of the Electrocochleogram in Clinical AudiometryInternational Journal of Audiology, 1973
- Human Acoustic Nerve Action Potential Recordings from the Tympanic Membrane Without AnesthesiaActa Oto-Laryngologica, 1972