Experience With Normal Breathhold Planning Scans for Radiosurgery of Moving Targets With Live Tracking
Authors:
- Jimm GrimmJGJimm Grimm
- Shiva F. NaidooSNShiva F. Naidoo
- Kristena YossefKYKristena Yossef
- Gaurav ShuklaGSGaurav Shukla
- Carla J. ScofieldCSCarla J. Scofield
- Abby C. SearfossASAbby C. Searfoss
- Madison A ZulkoskiMZMadison A Zulkoski
- James A TanyiJTJames A Tanyi
- Heath B. MackleyHMHeath B. Mackley
- Anand MahadevanAMAnand Mahadevan
Open Access
- 24 October 2022
- journal article
- research article
- Published by Springer Nature in Cureus
- Vol. 14 (10) , e30676
Abstract
Purpose: Utilization of breathhold scans with live tracking has a long track record of good published outcomes for stereotactic body radiation therapy (SBRT) and is recommended by the manufacturer of the Synchrony tracking system. However, the popularity of four-dimensional computed tomography (4DCT) scans challenges the validity of the breathhold scan with live tracking technique. Although this study is not intended to prove the superiority of either method, we demonstrate the feasibility of using the breathhold scans with a phantom test and clinical examples. Methods: A 4DCT of a perfect sphere was scanned at 20 breaths per minute and compared to a 4DCT of a small lung tumor in one patient and a 4DCT of a larger renal tumor in another patient, as well as to fiducial matching in a patient with pancreatic cancer. Normal exhale and normal inhale breathhold CT scans were performed for the pancreatic cancer patient, combined with Synchrony tracking on CyberKnife (Sunnyvale, CA: Accuray) for treatment. Results: The 4DCT scan of the phantom exhibited considerable apparent deformation, which must be entirely due to imaging artifact since the perfect sphere in the phantom is known to be completely rigid. The 4DCT of the lung and renal tumors in patients had similar apparent deformation. Usually in patients, from 4DCT alone, it is difficult to determine how much was due to deformation and how much was due to artifact. Fiducial positions in the final normal exhale and normal inhale breathhold scans for Synchrony matched each other within 1mm for the pancreatic cancer patient. Conclusion: We demonstrated the feasibility of breathhold scans with Synchrony live tracking, as recommended by the manufacturer. More studies will be needed to determine whether this method is better than using a 4DCT.Keywords
This publication has 28 references indexed in Scilit:
- Low toxicity for lung tumors near the mediastinum treated with stereotactic body radiation therapyPractical Radiation Oncology, 2013
- The Lessons of QUANTEC: Recommendations for Reporting and Gathering Data on Dose–Volume Dependencies of Treatment OutcomeInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Retrospective Analysis of Artifacts in Four-Dimensional CT Images of 50 Abdominal and Thoracic Radiotherapy PatientsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Intrafractional Motion of the Prostate During Hypofractionated RadiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Organ Deformation and Dose Coverage in Robotic Respiratory-Tracking RadiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Accuracy of tumor motion compensation algorithm from a robotic respiratory tracking system: A simulation studyMedical Physics, 2007
- Use of motion tracking in stereotactic body radiotherapy: Evaluation of uncertainty in off-target dose distribution and optimization strategiesActa Oncologica, 2006
- Feasibility report of image guided stereotactic body radiotherapy (IG-SBRT) with tomotherapy for early stage medically inoperable lung cancer using extreme hypofractionationActa Oncologica, 2006
- Four-dimensional CT scans for treatment planning in stereotactic radiotherapy for stage I lung cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004
- Respiration‐correlated spiral CT: A method of measuring respiratory‐induced anatomic motion for radiation treatment planningMedical Physics, 2002