Comparing the Efficacy and Safety of a Novel Monophasic Waveform Delivered by the Passive Implantable Atrial Defibrillator With Biphasic Waveforms in Cardioversion of Atrial Fibrillation
Open Access
- 6 April 2004
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 109 (13) , 1686-1692
Abstract
Background— The passive implantable atrial defibrillator (PIAD) (with no battery or discharging capacitor and powered transcutaneously by radio-frequency energy) delivering a novel monophasic low-tilt waveform is more efficacious than the standard monophasic waveform at atrial defibrillation. Standard biphasic (STB) waveforms, however, are more efficacious and safer than monophasic waveforms. This study compared the efficacy and safety of the PIAD waveform with biphasic waveforms. Methods and Results— Sustained atrial fibrillation (AF) was induced by rapid atrial pacing. Cardioversion was attempted via 2 atrial defibrillation leads. The efficacy of the PIAD was compared with 3 biphasic waveforms (standard, single rounded, and double rounded) at varying voltage settings in 10 pigs. After a synchronized shock, hemodynamic changes between the PIAD, standard biphasic, and monophasic waveforms were compared at 1.5 and 3.0 J in 12 pigs. Myocardial injury (biochemical and histological) after ten 5-J PIAD shocks was compared with a no-shock group in 14 pigs. The PIAD 100-V setting was significantly more efficacious than the STB (100/−50 V: 100% [1.88±0.02 J] versus 90% [0.89±0.0 J]; P =0.025). No arrhythmic, hemodynamic, or myocardial injury was observed with the PIAD waveform. Conclusions— Defibrillation with the PIAD is more efficacious than with the STB waveform and appears safe. This device could provide a more effective option for cardioversion.Keywords
This publication has 7 references indexed in Scilit:
- Novel Passive Implantable Atrial Defibrillator Using Transcutaneous Radiofrequency Energy Transmission Successfully Cardioverts Atrial FibrillationCirculation, 2003
- Waveform Optimization for Internal Atrial Defibrillation: Effects of Waveform Rounding, Phase Duration, and Voltage SwingPacing and Clinical Electrophysiology, 2001
- The effects of biphasic and conventional monophasic defibrillation on postresuscitation myocardial functionJACC, 1999
- Comparison of Cardiac Troponin I, Creatine Kinase-MB, and Myoglobin for Detection of Acute Ischemic Myocardial Injury in a Swine ModelAmerican Journal of Clinical Pathology, 1998
- Rounded Biphasic Waveform Reduces Energy Requirements for Transvenous Catheter Cardioversion of Atrial Fibrillation and FlutterPacing and Clinical Electrophysiology, 1997
- Testing different biphasic waveforms and capacitances: Effect on atrial defibrillation threshold and pain perceptionJACC, 1996
- Metabolic Changes and Mitochondrial Dysfunction Early Following Transthoracic Countershock in DogsPacing and Clinical Electrophysiology, 1989