Error in laboratory reference limits as shown in a collaborative quality-assurance program.
Authors:
- P J ComptonPCP J Compton
- M C StuartMSM C Stuart
- L LazarusLLL Lazarus
Open Access
- 1 May 1986
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 32 (5) , 845-849
Abstract
Laboratories participate in collaborative quality-assurance programs to maintain and improve the quality of their diagnostic assays, but little attention has been paid to diagnostic quality in these programs. We used a national quality-assurance program to assess the quality of laboratory reference intervals as exemplified by triiodothyronine, thyroxin, and thyrotropin immunoassays. The limits of the reference intervals used by laboratories bear virtually no relationship to laboratory bias, i.e., whether assays read "high" or "low." Further, correcting assay results from different laboratories for the reference limits used increases rather than decreases interlaboratory scatter. We conclude that the analytical quality of immunoassays now exceeds the quality of the reference limits supplied to clinicians to assist diagnosis, and that nationally or internationally defined reference limits would therefore provide more information at less cost than do individual laboratory reference limits.Keywords
This publication has 5 references indexed in Scilit:
- The Evaluation of Imprecision in Collaborative Immunoassay Quality-Assessment ProgrammesAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1984
- Progress report on a national quality-control survey of triiodothyronine and thyroxin assay.Clinical Chemistry, 1984
- An inter- and intralaboratory quality-control survey of radioimmunoassay of insulin, thyroxin, thyrotropin, cortisol, digoxin, gastrin, beta 2-microglobulin, and IgE in Japan with commercially available kits.Clinical Chemistry, 1983
- III. Application of Principles of Test Selection and InterpretationAnnals of Internal Medicine, 1981
- Quality Control of Radioimmunoassays for Proteins: The First Two and a Half Years of a National Scheme for Serum Growth Hormone MeasurementsAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1979