Reproducibility of Proton Magnetic Resonance Spectroscopy Imaging Measurements of Normal Human Hippocampus at 1.5 T: Clinical Implications
- 1 April 2001
- journal article
- Published by Wiley in Journal of Neuroimaging
- Vol. 11 (2) , 194-201
Abstract
The authors investigate the reproducibility of metabolite signals measured with proton magnetic resonance spectroscopy1H‐MRS) acquired from the human hippocampus in controls and in a phantom. Two1H‐MRS studies separated by 3 weeks were performed in 8 healthy volunteers and in a phantom. N‐acetyl compounds (NA), choline (Ch), and creatine (Cr) peak areas and ratios were measured and compared using percentage variation, and Pearson Correlation Coefficient at the level of every voxel, the level of 1 hippocampus (5 voxels), and the level of 2 hippocampi (10 voxels). Sensitivity for observing clinically significant between‐session1H‐MRS changes was evaluated using the reliable change index. Reproducibility measures for metabolite peak areas were only moderately concordant with percentage variation ranging from 14% to 20% for NA, Cho, and Cr. Stability was much improved when NA ratios and sum of multiple voxels were considered. Between‐session NA/(Cho + Cr) changes greater than 22%, 12%, and 10% in one given participant can be detected with a 90% confidence interval when considered at the single‐voxel level, the level of a single hippocampus, or the level of both hippocampi, respectively. Left‐right asymmetry indices showed similar and limited interhemispheric asymmetry in repeated examination. This study suggests that1H‐MRS reproducibility performance is adequate for the study and monitoring of human hippocampus function when NA ratios and the sum of multiple voxels are considered. Individual metabolite peaks and single‐voxel measurements have low reproducibility at 1.5 T and should be used only with clearly established statistical parameters.Keywords
This publication has 25 references indexed in Scilit:
- Quantitative 1H MRS of the human brain in vivo based on the simulation phantom calibration strategyMagnetic Resonance in Medicine, 1998
- Proton magnetic resonance spectroscopic imaging and magnetic resonance imaging volumetry in the lateralization of temporal lobe epilepsy: A series of 100 patientsAnnals of Neurology, 1997
- Imaging and epilepsyBrain, 1997
- Proton magnetic resonance spectroscopic imaging for discrimination of absence and complex partial seizuresAnnals of Neurology, 1997
- Empirical methods for assessing meaningful neuropsychological change following epilepsy surgeryJournal of the International Neuropsychological Society, 1996
- Normalization of contralateral metabolic function following temporal lobectomy demonstrated by 1H magnetic resonance spectroscopic imagingAnnals of Neurology, 1996
- Reproducibility of high spatial resolution proton magnetic resonance spectroscopic imaging in the human brainMagnetic Resonance in Medicine, 1996
- Lateralization of temporal lobe epilepsy based on regional metabolic abnormalities in proton magnetic resonance spectroscopic imagesAnnals of Neurology, 1994
- A method of assessing change in a single subject: An alteration of the RC indexBehavior Therapy, 1986
- Image-selected in Vivo spectroscopy (ISIS). A new technique for spatially selective nmr spectroscopyJournal of Magnetic Resonance (1969), 1986