Low–flow anaesthesia does not increase the risk of microbial contamination through the circle absorber system
Authors:
- 1 January 1989
- journal article
- research article
- Published by Wiley in Acta Anaesthesiologica Scandinavica
- Vol. 33 (1) , 89-92
Abstract
In the circle absorber system, a decrease in fresh gas flow means a higher degree of rebreathing, and, consequently, a higher temperature and humidity within the system. With our present hygienic routines, the circle system tubings are changed and decontaminated once daily. Thus, the same circle system is used for several patients each day. In order to evaluate whether the risk of bacterial contamination increased with the introduction of low–flow anaesthesia, 122 patients anaesthetized with either a low–flow technique (< 1.5 1 fresh gas flow/min) or with medium fresh gas flows (3–6 1/min) were studied. Bacterial samples were taken preoperatively from the oropharynx and postoperatively from five locations in the circle system. The patients were studied postoperatively for signs of respiratory tract infection. There were few positive bacteria cultures from the tubings in the circle system, regardless of fresh gas flow. No pathogens were found in the inspiratory tubings and no cases of postoperative respiratory tract infection could be related to bacterial spread from the anaesthesia machine. There were no indications that the present hygienic management was insufficient for the medium– or the low–flow circle system techniques.Keywords
This publication has 17 references indexed in Scilit:
- Preservation of humidity and heat of respiratory gases during anaesthesia ‐ a laboratory investigationActa Anaesthesiologica Scandinavica, 1987
- Colonisation of the trachea in ventilated patients: What is the bacterial pathway?Journal of Hospital Infection, 1983
- Bacterial Interactions between Anesthesiologists, Their Patients, and EquipmentAnesthesiology, 1982
- Contamination rate of sterilized ventilators in an ICUCritical Care Medicine, 1981
- Sterile Anesthesia Breathing Circuits Do Not Prevent Postoperative Pulmonary InfectionAnesthesiology, 1981
- Failure of Bacterial Filters to Reduce the Incidence of Pneumonia after Inhalation AnesthesiaAnesthesiology, 1981
- Lower respiratory tract infections associated with respiratory therapy and anaesthesia equipmentJournal of Hospital Infection, 1981
- The Anesthesia Machine and Circle System Are Not Likely to be Sources of Bacterial ContaminationAnesthesiology, 1977
- RETROGRADE SPREAD OF ORGANISMS FROM VENTILATOR TO PATIENT VIA THE EXPIRATORY LIMBThe Lancet, 1971