Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases
Open Access
- 11 June 2021
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 94 (1123) , 20210473
Abstract
Objectives: The aim of this work was to evaluate the operation of the 1600SRS detector and to develop a calibration procedure for verifying the dose delivered by a single isocenter stereotactic radiosurgery (SRS) treatment of small multiple brain metastases (BM). Methods: 14 clinical treatment cases were selected with the number of BM ranging from 2 to 11. The dosimetric agreement was investigated between the calculated and the measured dose by an OCTAVIUS 1600SRS array detector in an OCTAVIUS 4D phantom equipped with dedicated SRS top. The cross-calibration procedure deviated from the manufacturer’s as it applied field sizes and dose rates corresponding to the volumetric modulated arc therapy segments in each plan. Results: Measurements with a plan specific cross-calibration showed mean ± standard deviation (SD) agreement scores for cut-off values 50%, 80%, 95%, of 98.6 ± 1.7%, 96.5 ± 4.6%, 97.3 ± 4.4% for the 6 MV plans respectively, and 98.6 ± 1.5%, 96.6 ± 4.0% 96.4 ± 6.3%, for the 6 MV flattening filter free (FFF) plans respectively. Using the default calibration procedure instead of the plan specific calibration could lead to a combined systematic dose offset of 4.1% for our treatment plans. Conclusion: The 1600SRS detector array with the 4D phantom offers an accurate solution to perform routine quality assurance measurements of single isocenter SRS treatments of multiple BM. This work points out the necessity of an adapted cross-calibration procedure. Advances in knowledge: A dedicated calibration procedure enables accurate dosimetry with the 1600SRS detector for small field single isocenter SRS treatment of multiple brain metastases for a large amount of BM.Keywords
This publication has 25 references indexed in Scilit:
- Investigating ion recombination effects in a liquid-filled ionization chamber array used for IMRT QA measurementsMedical Physics, 2016
- Single-Isocenter Multiple-Target Stereotactic Radiosurgery: Risk of Compromised CoverageInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2015
- Defining the Optimal Planning Target Volume in Image-Guided Stereotactic Radiosurgery of Brain Metastases: Results of a Randomized TrialInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2015
- Performance parameters of a liquid filled ionization chamber arrayMedical Physics, 2013
- Plan quality and treatment planning technique for single isocenter cranial radiosurgery with volumetric modulated arc therapyPractical Radiation Oncology, 2012
- Consideration of optimal isodose surface selection for target coverage in micro-multileaf collimator-based stereotactic radiotherapy for large cystic brain metastases: comparison of 90%, 80% and 70% isodose surface-based planningBritish Journal of Radiology, 2012
- Inter- and Intrafraction Patient Positioning Uncertainties for Intracranial Radiotherapy: A Study of Four Frameless, Thermoplastic Mask-Based Immobilization Strategies Using Daily Cone-Beam CTInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm: fundamental characterization in waterPhysics in Medicine & Biology, 2011
- Feasibility of Single-Isocenter Volumetric Modulated Arc Radiosurgery for Treatment of Multiple Brain MetastasesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Comparison of dose calculation algorithms for treatment planning in external photon beam therapy for clinical situationsPhysics in Medicine & Biology, 2006