Optimization of a Procedure Used to Measure Aerosol Characteristics of Nebulized Solutions Using a Cooled Next Generation Impactor
Authors:
- 1 December 2010
- journal article
- research article
- Published by SAGE Publications in Journal of Aerosol Medicine and Pulmonary Drug Delivery
- Vol. 23 (6) , 397-404
Abstract
Background: Cooling the Next Generation Impactor (NGI) is recommended to minimize evaporation due to heat transfer from impactor to aerosols when evaluating nebulized solutions. This methodology increases testing time for serial testing procedures. We hypothesize that after an initial prolonged cooling time, experiments could be repeated after shorter recooling times without sacrificing accuracy. Methods: Three units of continuous output (HUDSON) and breath enhanced (PARI LC Plus) nebulizers were operated (6 L/min) with albuterol solution (2.5 mg/3 mL) into a cooled (4°C) NGI (internal and external filters) calibrated at 15 L/min. Mass median aerodynamic diameter (MMAD), geometric standard deviation (GSD), % particles Results: MMAD, GSD, P% 0.05). Percentage of variation from first measurement ranged from: −3.9 to +2.1% (PARI) and −4.1 to +2.9% (HUDSON) for MMAD; −5.6 to +2.6% (PARI) and −4.9 to +1.9% (HUDSON) for GSD; 0 to +4.6% (PARI) and −3.7% to +5.7% (HUDSON) for P%Conclusions: Aerosol characteristics of nebulized solutions determined by NGI are not affected by performing two repeat measurements after recooling the impactor for either 30 or 60 min after an initial 90-min time.Keywords
This publication has 6 references indexed in Scilit:
- Evolution of Evaporative Understanding within Nebulizer StandardsJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2009
- Aerodynamic Characteristics of Nebulized Terbutaline Sulphate Using the Next Generation Impactor (NGI) and CEN MethodJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2009
- Determination of Nebulizer Droplet Size Distribution: A Method Based on Impactor RefrigerationJournal of Aerosol Medicine, 2007
- Next Generation Pharmaceutical Impactor: A New Impactor for Pharmaceutical Inhaler Testing. Part III. Extension of Archival Calibration to 15 L/minJournal of Aerosol Medicine, 2004
- Undersizing of droplets from a vented nebulizer caused by aerosol heating during transit through an anderson impactorJournal of Aerosol Science, 1999
- Four Hours of Continuous Albuterol NebulizationChest, 1998