Abstract
Background: Cooling the Next Generation Impactor (NGI) is recommended to minimize evaporation due to heat transfer from impactor to aerosols when evaluating nebulized solutions. This methodology increases testing time for serial testing procedures. We hypothesize that after an initial prolonged cooling time, experiments could be repeated after shorter recooling times without sacrificing accuracy. Methods: Three units of continuous output (HUDSON) and breath enhanced (PARI LC Plus) nebulizers were operated (6 L/min) with albuterol solution (2.5 mg/3 mL) into a cooled (4°C) NGI (internal and external filters) calibrated at 15 L/min. Mass median aerodynamic diameter (MMAD), geometric standard deviation (GSD), % particles Results: MMAD, GSD, P% 0.05). Percentage of variation from first measurement ranged from: −3.9 to +2.1% (PARI) and −4.1 to +2.9% (HUDSON) for MMAD; −5.6 to +2.6% (PARI) and −4.9 to +1.9% (HUDSON) for GSD; 0 to +4.6% (PARI) and −3.7% to +5.7% (HUDSON) for P%Conclusions: Aerosol characteristics of nebulized solutions determined by NGI are not affected by performing two repeat measurements after recooling the impactor for either 30 or 60 min after an initial 90-min time.