A thin glycoprotein coating of a synthetic lenticule does not cause nutritional deficiency of the anterior cornea
Authors:
- Deborah F. SweeneyDSDeborah F. Sweeney
- Hans J. GriesserHGHans J. Griesser
- Simon ToutSTSimon Tout
- Han Ying ChengHCHan Ying Cheng
- John G. SteeleJSJohn G. Steele
- 1 January 1999
- journal article
- Published by Taylor & Francis in Current Eye Research
- Vol. 18 (5) , 335-341
Abstract
Purpose. This study investigated whether a glycoprotein coating that will be used to enhance corneal epithelialization affects in situ nutritional passage through a permeable membrane. Methods. Sixteen adult cats were equally divided into two groups. Polycarbonate membranes with pore size of 0.1æm and total pore area (porosity) of 3.1% were used as implant materials. The membranes for Group 1 were coated with a thin layer of Collagen I, while the membranes for Group 2 were uncoated. Each membrane with 8-mm diameter was implanted into an interlamellar pocket of the cornea. The eyes were observed for approximately 35 days to monitor clinical signs of nutritional deficiency of the cornea, and then 7 membranes were removed from the eyes. The permeability of the explanted membranes to glucose, inulin and albumin was used to predict the in situ difference between the coated and uncoated groups in regard to nutritional passage through the membranes. To investigate the long-term effect of the surface coating on corneal health, two animals from Group 1 were followed for up to two years and then both eyes of each animal underwent histological examination. Results. Clinically, no post-surgical complications associated with nutritional deficiency were observed in any of the eyes. Nutritional permeability tests showed no significant differences between the coated and uncoated membranes. Histologically, the long-term animals showed no abnormal morphology associated with nutritional deficiency in the cornea anterior or posterior to the membranes. Conclusions. A thin glycoprotein coating on a permeable membrane does not appear to affect the nutritional supply of the anterior cornea and therefore can be used to enhance epithelialization of synthetic corneal onlays in vivo.Keywords
This publication has 7 references indexed in Scilit:
- Surface topography can interfere with epithelial tissue migrationJournal of Biomedical Materials Research, 1998
- Refractive Keratoplasty with Synthetic Lens ImplantsInternational Ophthalmology Clinics, 1991
- Corneal cell adhesion and proliferation on hydrogel sheets bound with cell-adhesive proteinsCurrent Eye Research, 1991
- Modeling Glucose Distribution in the CorneaCurrent Eye Research, 1990
- Polysulfone corneal lensesJournal of Cataract & Refractive Surgery, 1986
- Basement membrane components in healing rabbit corneal epithelial wounds: immunofluorescence and ultrastructural studies.Journal of Cell Biology, 1984
- The Effect of Intralamellar Water-Impermeable Membranes on Corneal HydrationArchives of Ophthalmology (1950), 1966