In Vitro and In Vivo Determination of the Thermal Effect of Middle Ear Endoscopy
Authors:
- Donald F. PerraultDPDonald F. Perrault
- 1 February 1996
- journal article
- research article
- Published by Wiley in The Laryngoscope
- Vol. 106 (2) , 213-216
Abstract
During middle ear endoscopy, patients frequently experience vertigo. We theorized that this effect is secondary to heat produced by the scope. We evaluated two Hopkins rods and two fiberoptic scopes to assess the temperature elevation at the lateral semicircular canal (LSCC) of temporal bones and in a live canine model. These results were compared to the results of a standard 44° C water caloric.We demonstrated in both models that the temperature rise at the LSCC increased relative to the scope diameter and that the Hopkins rods produced the same or greater heating effect than a warm caloric. Direct exposure of the thermocouple to the light produced a greater temperature elevation.We conclude that endoscopes produce sufficient heat to induce caloric stimulation of the LSCC. Care should be exercised to prevent the possibility of thermal injury.Keywords
This publication has 4 references indexed in Scilit:
- Transtympanic endoscopy of the middle earOperative Techniques in Otolaryngology-Head and Neck Surgery, 1992
- Transtympanic endoscopy of the middle earThe Laryngoscope, 1992
- Fiberoptic Endotoscopes for Examining the Middle EarJAMA Otolaryngology–Head & Neck Surgery, 1967
- Temperature Changes in the Perilymph Space in Response to Caloric Stimulation in ManActa Oto-Laryngologica, 1954