Clinical evaluation of a high-resolution new peripheral quantitative computerized tomography (pQCT) scanner for the bone densitometry at the lower limbs
Authors:
- 1 August 1998
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 43 (8) , 2279-2294
Abstract
Precision, long-term stability, linearity and accuracy of the x-ray peripheral quantitative computerized tomographic (pQCT) bone scanner XCT 3000 (Norland-Stratec Medical Sys.) were evaluated using the European Forearm Phantom (EFP). In vivo measurements were assessed using a standardized procedure at the distal femur and the distal tibia. In the patient-scan mode, the spatial resolution of the system was 1.04 +/- 0.05 lp/mm as measured at the 10% level of the modulation transfer function (MTF). The contrast-detail diagram (CDD) yielded a minimal difference in attenuation coefficient (AC) of 0.07 cm(-1) at an object size of 0.5 mm. The effective dose for humans was calculated to be less than 1.5 microSv per scan. Short-term precision in vivo was expressed as root mean square standard deviation of paired measurements of 20 healthy volunteers (RMSSD = 0.5%). At the distal femur total volumetric density (ToD) and total cross-sectional area (ToA) were found to be less sensitive to positioning errors than at the distal tibia. Structural parameters like the polar cross-sectional moment of inertia (CSMIp) or the polar cross-sectional moment of resistance (CSMRp) showed a good short-term precision at the distal femur (RMSSD = 1.2 and 1.4%). The relation between the two skeletal sites with respect to CSMIp or CSMRp showed a high coefficient of determination (r2 = 0.77 and 0.74).Keywords
This publication has 33 references indexed in Scilit:
- Heterogeneity of trabecular and cortical postmenopausal bone loss: A longitudinal study with pQCTBone, 1997
- Appendicular measurements in screening women for low axial bone mineral densityBritish Journal of Radiology, 1996
- Dexamethasone Effects on Mechanical, Geometric and Densitometric Properties of Rat Femur Diaphyses As Described by Peripheral Quantitative Computerized Tomography and Bending TestsBone, 1995
- Stellenwert zweier unterschiedlicher Knochendichtemeßmethoden zur Bestimmung des Mineralgehalts am peripheren und axialen SkelettZeitschrift für Orthopädie und ihre Grenzgebiete, 1992
- Abschätzung der effektiven Dosis bei Knochenmineralmessungen mit Photonenabsorptiometrie und ComputertomographieRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1991
- Differential effects of aging and disease on trabecular and compact bone density of the radiusBone, 1991
- Precision of measurement of bone density with a special purpose computed tomography scannerBritish Journal of Radiology, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- A gamma-ray computed tomography scanner for the quantitative measurement of bone densityJournal of Biomedical Engineering, 1985
- Kleincomputertomograph für quantitative Knochenuntersuchungen in den Extremitäten des Menschen - Special Purpose CT-System for Quantitative Bone Evaluations in the Appendicular SkeletonBiomedical Engineering / Biomedizinische Technik, 1983