Redilation of e‐PTFE covered CP stents
Authors:
- Gianfranco ButeraGBGianfranco Butera
- Gianpiero GaioGGGianpiero Gaio
- Mario CarminatiMCMario Carminati
- 11 April 2008
- journal article
- pediatric and-congenital-heart-disease
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 72 (2) , 273-277
Abstract
Objectives: To evaluate the possibility to redilate covered Cheatham‐Platinum stents during follow‐up, in particular in growing children with aortic coarctation. Background: There are no data in the literature about the redilation of ePTFE covered CP stents. Methods: Sixty covered CP stents were implanted in patients with aortic coarctation or recoarctation between January 2004 and October 2007. Seven patients (mean age 14.2 ± 3.7 years) needed to repeat the hemodynamic study due to somatic growth and increase of aortic gradient with the occurrence of systemic hypertension. Two had near‐atretic aortic coarctation, three had postsurgical recoarctation and aneurysm formation, one had native aortic coarctation associated with aneurysm of the arterial wall, and one had severe native aortic coarctation. Results: Procedures were performed a mean of 20 ± 5 months (range, 12–24 months) after the primary stent implantation. Fluoroscopy time ranged between 7 and 15 min (median, 10 min) whereas procedure time ranged between 60 and 75 min (median, 65 min). After redilation the gradient across the stenosis decreased from a median value of 35 mm Hg to a median value of 5 mm Hg. The stent diameter increased of 20–50% the predilation value. No complications occurred and angiographic controls showed that the stenoses have been relieved. Follow‐up: During a median follow‐up of 12 months (6–30 months) the results were stable without complications. Conclusion: Covered Cheatham‐Platinum stents can be easily redilatedKeywords
This publication has 15 references indexed in Scilit:
- Closure of extracardiac Fontan fenestration by using the covered Cheatham Platinum stentCatheterization and Cardiovascular Interventions, 2007
- Aortic rupture following a covered stent for coarctation: Delayed recognitionCatheterization and Cardiovascular Interventions, 2006
- Covered Cheatham-Platinum Stents for Aortic CoarctationJACC, 2006
- Covered stents in patients with congenital heart defectsCatheterization and Cardiovascular Interventions, 2006
- The CP stent--short, long, covered--for the treatment of aortic coarctation, stenosis of pulmonary arteries and caval veins, and Fontan anastomosis in children and adults: an evaluation of 60 stents in 53 patientsHeart, 2005
- Use of Covered Stents in the Management of Coarctation of the AortaPediatric Cardiology, 2004
- Long-term outcome (up to 15 years) of balloon angioplasty of discrete native coarctation of the aorta in adolescents and adultsJACC, 2004
- Transcatheter Treatment of Coarctation of the Aorta: A ReviewPediatric Cardiology, 1998
- Exercise Blood Pressure Response and 5-Year Risk of Elevated Blood Pressure in a Cohort of Young Adults: The CARDIA StudyAmerican Journal of Hypertension, 1994
- Is There a Relationship between Exercise Systolic Blood Pressure Response and Left Ventricular Mass?Annals of Internal Medicine, 1992