Magnetic Field Measurement Using Peak-Locked Zeeman Coherent Population Trapping Resonance in Rubidium Vapor
Authors:
Open Access
- 13 June 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Photonics Journal
- Vol. 11 (4) , 1-10
Abstract
A sensor for magnetic field measurement is developed using a small (centimeter-size) rubidium cell. High contrast Zeeman coherent population trapping (CPT) resonance is produced by the magnetic dither in the sensor solenoid, and a peak-locked servo is implemented for magnetic field measurement. Sensitivity close to $100\ pT / \sqrt{Hz}$ has been achieved in axial field measurement. The CPT based magnetic sensor is capable of measuring sub-nanotesla level magnetic field variations at high (microtesla) field strength. We discuss possible sources of noise and broadening which are affecting the performance of the sensor, and suggest changes that will further improve the performance of the sensor.
Keywords
Funding Information
- NASA-EPSCoR (80NSSC17M0026)
- NASA-MIRO (NNX15AP84A)
- DOD (62818-PH-REP)
This publication has 29 references indexed in Scilit:
- 85Rb Coherent-Population-Trapping Atomic Clock with Low Power ConsumptionJapanese Journal of Applied Physics, 2009
- Slow Light and Its ApplicationsJournal of the Optical Society of America B, 2008
- Ultrahigh enhancement in absolute and relative rotation sensing using fast and slow lightPhysical Review A, 2007
- Nonlinear magneto-optical rotation with frequency-modulated light in the geophysical field rangePhysical Review A, 2006
- Push-Pull Optical Pumping of Pure Superposition StatesPhysical Review Letters, 2004
- Recent Progress in Optically Pumped MagnetometersPhysica Scripta, 2003
- Experimental realization of coherent dark-state magnetometersEurophysics Letters, 1998
- Giant Kerr nonlinearities obtained by electromagnetically induced transparencyOptics Letters, 1996
- Steep dispersion and group velocity belowin coherent population trappingPhysical Review A, 1996
- Quantum sensitivity limits of an optical magnetometer based on atomic phase coherencePhysical Review A, 1994