Effect of tear osmolarity on repeatability of keratometry for cataract surgery planning
- 1 August 2015
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 41 (8) , 1672-1677
Abstract
To evaluate the effects of tear osmolarity on the repeatability of keratometry (K) measurements in patients presenting for cataract surgery. Three clinical practices. Observational prospective nonrandomized study. Subjects were prospectively recruited based on tear osmolarity (Tearlab Osmolarity System); that is, osmolarity more than 316 mOsm/L in at least 1 eye (hyperosmolar) and osmolarity less than 308 mOsm/L in both eyes (normal). The baseline K value was measured, and a second measurement was taken on the same instrument (IOLMaster) within 3 weeks of the first. Variability in average K, calculated corneal astigmatism using vector analysis, and intraocular lens (IOL) sphere power calculations were compared between groups. The hyperosmolar group (50 subjects) had a statistically significantly higher variability in the average K reading (P = .05) than the normal group (25 subjects) and a statistically significantly higher percentage of eyes with a 1.0 diopter (D) or greater difference in the measured corneal astigmatism (P = .02). A statistically significantly higher percentage of eyes in the hyperosmolar group had an IOL power difference of more than 0.5 D (P = .02). No statistically significant differences were present when the subjects were grouped by self-reported dry eye. Significantly more variability in average K and anterior corneal astigmatism was observed in the hyperosmolar group, with significant resultant differences in IOL power calculations. Variability was not significantly different when subjects were grouped by self-reported dry eye. Measurement of tear osmolarity at the time of cataract surgery planning can effectively identify patients with a higher likelihood of high unexpected refractive error resulting from inaccurate keratometry. Drs. Epitropoulos, Matossian, Berdy, and Malhotra received compensation from Tearlab for participating in the study. No author has a financial or proprietary interest in any material or method mentioned.Keywords
This publication has 19 references indexed in Scilit:
- Comparison of ocular biometry and intraocular lens power using a new biometer and a standard biometerJournal of Cataract & Refractive Surgery, 2014
- Investigation of Tear Osmolarity Using the TearLab Osmolarity System in Normal Adults in Saudi ArabiaEye & Contact Lens: Science & Clinical Practice, 2014
- Comparability and repeatability of corneal astigmatism measurements using different measurement technologiesJournal of Cataract & Refractive Surgery, 2012
- Top 5 Pearls to Consider When Implanting Advanced-technology IOLs in Patients With Ocular Surface DiseaseInternational Ophthalmology Clinics, 2012
- Tear Osmolarity in the Diagnosis and Management of Dry Eye DiseaseAmerican Journal of Ophthalmology, 2011
- Comparison of the ultrasonographic method with 2 partial coherence interferometry methods for intraocular lens power calculationOptometry - Journal of the American Optometric Association, 2011
- Preoperative Evaluation of Patients Before Cataract and Refractive SurgeryInternational Ophthalmology Clinics, 2011
- Precision of biometry, keratometry, and refractive measurements with a partial coherence interferometry–keratometry deviceJournal of Cataract & Refractive Surgery, 2010
- Hyperosmolarity of the Tear Film in Dry Eye Syndrome2010
- Comparison of portable automated keratometry and manual keratometry for IOL calculationJournal of Cataract & Refractive Surgery, 1997