Calibration of an ultra-low-background proportional counter for measuring [sup 37]Ar
- 12 August 2013
- conference paper
- Published by AIP Publishing in AIP Conference Proceedings
- Vol. 1549 (1) , 26-29
Abstract
An ultra-low-background proportional counter design has been developed at Pacific Northwest National Laboratory (PNNL) using clean materials, primarily electro-chemically-purified copper. This detector, along with an ultra-low-background counting system (ULBCS), was developed to complement a new shallow underground laboratory (30 meters water-equivalent) at PNNL. The ULBCS design includes passive neutron and gamma shielding, along with an active cosmic-veto system. This system provides a capability for making ultra-sensitive measurements to support applications like age-dating soil hydrocarbons with , age-dating of groundwater with , and soil-gas assay for to support On-Site Inspection (OSI). On-Site Inspection is a key component of the verification regime for the Comprehensive Nuclear-Test-Ban Treaty (CTBT). Measurements of radionuclides created by an underground nuclear explosion are valuable signatures of a Treaty violation. For OSI, the 35-day half-life of , produced from neutron interactions with calcium in soil, provides both high specific activity and sufficient time for inspection before decay limits sensitivity. This work describes the calibration techniques and analysis methods developed to enable quantitative measurements of samples over a broad range of proportional counter operating pressures. These efforts, along with parallel work in progress on gas chemistry separation, are expected to provide a significant new capability for soil gas background studies.